Herunterladen Inhalt Inhalt Diese Seite drucken

HBM TIM-PN Montageanleitung Seite 55

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Supporting points
approximation
Input interpolation
equation
Number of
measurement points
Rotor ID No.
TIM-PN
Off
Torque low-pass filter 1
Torque low-pass filter 2
Torque low-pass filters 1 and 2
On; Off
2; 3; 5; 7; 9; 11
N/A
Example 1
Input of of interpolation points
Select the channel to which Supporting points approxima­
tion should be applied. Now select Table from the Sup­
porting points approximation menu item and then se­
lect the number of measurement points.
Enter the setpoint values and actual values in the table.
You must enter the values for both the clockwise torque
and the anticlockwise torque.
In two-point scaling (e.g. using the manufacturing certific­
ate, working standard calibration), select the number 2
measurement points from the table and enter the corres­
ponding values for zero point and span.
After the Save button is pressed, the gradient factor for
clockwise and anticlockwise torque is calculated and dis­
played in the table. Then activate Supporting points
approximation for low-pass filter 1 and/or 2 with ON, or
deactivate this with OFF.
Clockwise torque = positive torque
Anticlockwise torque = negative torque
A4341-2.0 HBM: public
Settings
53