Herunterladen Inhalt Inhalt Diese Seite drucken

HBM TIM-PN Montageanleitung Seite 52

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Settings
Speed system cannot
be sufficiently
determined
[static]
Compatibility failure
[static]
Speed system not
present [static]
Speed system cannot
be sufficiently
identified
[static]
Reset with zero index
[static]
Compatibility failure
[static]
50
Web browser selection is set to "determine automatically from
transducer" and the torque transducer cannot determine the
properties Pulses/Rev., Max. Speed, Type.
Or the selection is set to "Manual" and the maximum speed is
set to 0 or "-".
The stator and rotor speed systems are not compatible in
physics or in version.
Table of failures in angle of rotation
Web browser selection is set to "determine automatically from
transducer" and the torque transducer is not outputting any
information about the speed system or a speed system is not
present.
Web browser selection is set to "determine automatically from
transducer" and the torque transducer cannot determine the
properties Pulses/Rev., Max. Speed, Type.
The web browser selection is set to "determine automatically
from transducer" and reset with zero index active, but no zero
index is present or cannot be determined if present
The stator and rotor speed systems are not compatible in
physics or in version.
A4341-2.0 HBM: public
TIM-PN

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis