HBM TIM-PN Montageanleitung Seite 51

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

CRC5 failure in
sequence
[dynamic]
Measured value not
sent (stator has
detected failures or
transmission is
defective)
[dynamic]
Linearization failure
[static]
Measured value
overflow or
underflow
[dynamic]
Speed system not
present [static]
Speed value overflow
or underflow
[dynamic]
TIM-PN
Filter
1 Hz
0.1 Hz
Table of errors in torque
During transmission from the torque transducer to TIM-PN, a
CRC error is detected in more than 5 values in direct
sequence (otherwise the preceding, old, correct value is
taken)
If a measured value is not sent at all, the failure state must
be assumed immediately. If the stator can be accessed, the
device failures are also updated.
Linearization is active, but the rotor serial number is different
to that in the TIM-PN device.
The limit for T40 is +/- 120% of the nominal (rated) torque.
Table of failures in rotational speed
Web browser selection is set to "determine automatically from
transducer" and the torque transducer is not outputting any
information about the speed system or a speed system is not
present.
The speed value exceeds the maximum value (nominal
(rated) rotational speed + 105%)
A4341-2.0 HBM: public
Failure flag duration on PROFINET
975 ms
27 s
Settings
49