10.3.5 Supporting points approximation

Fig. 10.15 Supporting points approximation
The Supporting points approximation menu can be used
to re-adjust the torque. This can be done either by enter­
ing the measured values (interpolation points) or by en­
tering the gradient. The approximation is only valid for the
current rotor (the rotor ID number must match the con­
nected rotor). The rotor ID is shown on the start page.
The Supporting points approximation parameters are
saved in the TIM-PN. These data are not stored in the
parameter set. If the entered rotor ID does not match the
connected rotor, then Supporting points approximation
will be deactivated.