TIM-PN
Information
If settings are changed in the web browser, this has a
direct influence on the measurement signal. If, for ex­
ample, you change the decimal places for the torque, the
data on the PROFINET interface is immediately sent with
the new decimal places.
As the data is transmitted in 32-bit integer format, this
can cause incorrect interpretation of the measurement
data by the PROFINET Master.
The standard settings are underlined.
E.g. Torque = 310.2Nm
Decimal places = 2 -> PROFINET value = 31020
Decimal places = 3 -> PROFINET value = 310200
Torque
Nm; kNm; ozfin; ozfft; lbfin; lbfft
.; .0; .00; .000; .0000; .00000
Positive; negative
0.1 Hz ; 1 Hz; 10 Hz; 100 Hz; 1 kHz; 3 kHz; CASMA filter;
Off;
0.1 Hz ; 1 Hz; 10 Hz; 100 Hz; 1 kHz; 3 kHz; Off
Speed
rpm ; rps
.; .0; .00; .000
Positive ; negative