Unit
Decimal point
Sign
Low-pass filter 1
(-1 dB)
Low-pass filter 2
(-1 dB)
Unit
Decimal point
Sign
TIM-PN
Information
If settings are changed in the web browser, this has a
direct influence on the measurement signal. If, for ex
ample, you change the decimal places for the torque, the
data on the PROFINET interface is immediately sent with
the new decimal places.
As the data is transmitted in 32-bit integer format, this
can cause incorrect interpretation of the measurement
data by the PROFINET Master.
The standard settings are underlined.
E.g. Torque = 310.2Nm
Decimal places = 2 -> PROFINET value = 31020
Decimal places = 3 -> PROFINET value = 310200
Torque
Nm; kNm; ozfin; ozfft; lbfin; lbfft
.; .0; .00; .000; .0000; .00000
Positive; negative
0.1 Hz ; 1 Hz; 10 Hz; 100 Hz; 1 kHz; 3 kHz; CASMA filter;
Off;
0.1 Hz ; 1 Hz; 10 Hz; 100 Hz; 1 kHz; 3 kHz; Off
Speed
rpm ; rps
.; .0; .00; .000
Positive ; negative