TP1
TIM-PN
Information
The maximum number of measured values that can be
averaged is n
max
this value is not exceeded. The number of n can be re­
duced by using the divisor.
You must also ensure that the angle of rotation pulses do
not arrive any faster than the torque measurements. Oth­
erwise the calculation of the moving average will be in­
correct, as the applied torque measurement will be taken
into account more than once in the average value filter.
This means that the maximum permissible rotational
speed must be defined and taken into account in the act­
ive CASMA filter.
In the TIM-PN, the CASMA filter is only available in signal
path TP1. In signal path TP1, the filter can be switched
on ć or switched (OFF) to bypass mode.
CASMA filter
Fig. 10.11 Signal path
A4341-2.0 HBM: public
= 4096 values. So you must ensure that
ON/OFF
TP1: PROFINET/UDP
Settings
41