Adjust the setting scale
The adjustment of the pressure reducer is lost when the adjustment handle is removed.
Protective equipment:
Material:
Close the shut-off valve on the inlet side.
1.
Relieve the pressure on the output side, e.g. with a water tap.
2.
Close the shut-off valve on the outlet side.
3.
Install the pressure gauge.
4.
CAUTION!
Risk of injury due to the pressure spring jumping out
If the pressure spring jumps out of the spring cover, this can lead to injuries.
– Do not unscrew the slotted screw
– Release pressure spring
Unscrew the slotted screw.
5.
Slowly open the shut-off valve on the inlet side.
6.
Set the desired back pressure.
7.
Check the sealing disc, the nozzle edge and the groove ring for perfect condition and
8.
replace the complete valve insert if necessary.
Align the scale value with the mark in the middle of the window.
9.
10. Tighten the slotted screw.
11. Slowly open the shut-off valve on the outlet side.
Chemical-resistant protective gloves
n
Protective goggles
n
Safety shoes
n
Pressure gauge
n
61
Maintenance
MAN048969, Ver. 5-05.2024