Start-up
7.6

Set the dosage

NOTICE!
Do not use tools!
Dosing must not be set under pressure!
Fig. 10: Set the dosage
Prerequisite:
The station is connected to the water supply.
n
The pressure regulator is set to a maximum water pressure of 6 bar.
n
The dosing pump has been bled.
n
Open the safety gate.
1.
Close ( Fig. 9 ,
2.
Unfasten the tapered lock nut ( Fig. 10 ,
3.
Unfasten or tighten the setting nut
4.
dosing.
Tighten the lock nut, holding the setting nut in the desired position.
5.
MAN048969, Ver. 5-05.2024
3
2
1
) the water supply.
1
The prescribed dosage setting for Ecolab Inspexx 210 is 0,16%
Lock nut
1
Adjusting nut
2
Top edge of setting nut
3
).
1
until the top edge
2
3
44
displays the desired
± 0,02%.