An external grounding bolt (1) designed for equalpotential bonding is provided on the IND500x terminal.
Figure 52: Grounding Bolt

2.7 PCB Switch Settings

Four switches (SW1 and SW2, identified in [Main PCB   Page 9]) are located on the main PCB. Functions of
the switches are listed in the following table.
Main PCB Switches Function
SW1-1
SW1-2
OFF
OFF
OFF
ON
OFF
ON
OFF
OFF
OFF
OFF
ON
ON
ON
ON
When both SW1-1 and SW1-2 are positioned to ON and power is applied to the terminal, a Master Reset
function will be initiated. This procedure will erase all programming in the terminal and return all settings back
to factory default values. Metrologically significant data (calibration) is not reset unless SW2-1 is also set to
ON during this process. This process is further described in the IND500x Technical Manual, Chapter 5,
Maintenance and Service.

2.8 PCB Jumper Position

2.8.1 Enable or Disable the Power On/Off Key

The Power On/Off key is disabled by default.
W1 Jumper Description
State
Close
Open (default) Disables the Power On/Off key on the terminal keypad. Terminal will always power up
IND500x
1
SW2-1
SW2-2
Function
OFF
OFF
For normal operation.
OFF
OFF
For normal operation and prohibits changes to metrologically
significant areas in the menu tree.
OFF
OFF
For firmware update. When the four switches are set this way, the
terminal is then ready for firmware update. Refer to the IND500x's
Service Manual for instructions on firmware update.
OFF
For connection to the InSite
ON
port.
OFF
Factory test.
ON
OFF
OFF
Master Reset: Master reset of the terminal, but metrologically
significant data, such are scale calibration, GEO code, etc., will
not be reset.
ON
OFF
Factory Reset: Master reset of the terminal, including metrolog-
ically significant data.
Description
Enables the Power On/Off key on the terminal keypad. When Ex power supply is connected,
the Power On/Off key is pressed to start the terminal.
automatically after a power outage.
TM
CSL program using the COM1 serial
Installation​ ​
45