The Alibi memory operates by storing up to 600 Alibi records in a battery-backed file as they occur. After this
file is full, all of these records are written to the "alibi.bin" file in flash and the 600-record file is cleared and
begins to store the next 600 records. The flash file (alibi.bin) can store up to 100,000 transactions before it
rolls over and begins to overwrite the oldest file.
Each record in the Alibi Memory file includes:
Figure 61: Alibi Transaction Record
Date and time stamp fields
A transaction counter value, which is a unique numeric field that identifies the transaction (the transaction
counter must be enabled in terminal setup to activate the transaction counter value)
Gross or net weight, tare weight, preset tare, and weight unit

3.4.1.1 View Alibi Memory Records

Alibi memory records can be viewed in setup at Application > Memory > Alibi, or they can be accessed by
using either the REPORTS softkey
1 Press the REPORTS softkey
2 If the REPORTS softkey
TABLE softkey
3 If the Alibi softkey
screen shows directly.
Figure 62: Alibi
4 To filter the Alibi records, see View and Print Alibi Memory.
The Alibi Memory cannot be manually cleared, but automatically cleared after the function is disabled and
enabled again.

3.4.2 Tare Table

The IND500x terminal contains a tare table with 199 records for storing tare weights that can be recalled by the
operator for use instead of manually entering them for each transaction. This recall function is especially useful
when certain tare values are used repeatedly. When totalization is enabled for the tare table, each time a trans-
action is completed using an activated tare ID, the selected weight value (gross or net weight) is added to the
total and the counter increments by one.
For gross weight accumulation, the tare table can be used with tare values of 0 in order to accumulate weight
by tare ID.
The counter for the tare totals is seven-digit long and has a maximum value of 1,500,000. When this value is
exceeded, an overflow error displays and that value is not accumulated. The counter must be reset in order to
continue totalizing. The total register is 11 digits in length including all positions to the right of the decimal
point. The decimal position is determined by the display resolution for the unit entered as the tare unit. The
maximum value for a scale with a display resolution of 0.01 kg would be 999999999.99 kg. When this value
is exceeded, an overflow error displays and that value is not accumulated. The total must be reset in order to
continue totalizing.
These tare weights can be recalled by either selecting from a list of all the values using the TARE TABLE softkey
followed by the FILTER OFF softkey
followed by the TARE TABLE softkey
IND500x
or the Alibi softkey
or the Alibi softkey
is pressed, select Alibi Memory from the selection box shown. Press the VIEW
shown at the bottom of the screen. The Alibi screen displays.
is pressed instead of the REPORTS softkey
or can be "quick accessed" directly by pressing the Tare ID value
.
from the home screens.
.
from the Home screen, the Alibi