8-7. Output

52
To output measured data via the RS-232C, choose one of the following three methods.
The timing charts shown below are for EXT.IN1 and mini jack input. The high level and low
level logic are reversed for EXT.IN2 input.
1. Input to the EXT.IN terminal 1
Initial setting:
(set at time of shipping)
This mode is for when the PZ201 foot-switch (optional) is used. Data is output A ms
after the EXT.IN signal goes ON.
EXT. IN
A
2. Input to the EXT.IN terminal 2
Initial setting:
Mode for shortening the output interval. (Processed as an interrupt.)
However, there must be no chattering in the input signal. Data is output when the
EXT.IN signal goes ON.
EXT. IN
A'
Data signaling rate (bps)
MIN. A
19200
21
9600
21
4800
21
2400
21
1200
21
600
21
3. Commands
Data is output by input ASCII code commands. For more details, refer to section 8-8.
Note
• After the EXT.IN signal goes ON, the main display is locked to the status just before the
EXT.IN signal went ON until RS-232C data output finishes. In addition, front panel key
operations are also not accepted during this period.
• When the initial setting of counter board I/O connector pins 2!§ is
judgment (I/O connector output, front panel LED) is held while the EXT.IN signal is ON.
However, note that the Go/No Go judgment is held for only unit 1 during link connection.
B
B'
MIN. A'
1
2
3
4
20
30
40
4
30
47
67
4
47
84
121 158 195 232 269 306 343 380
4
83 156 229 302 375 448 521 594 667 740
4
115 300 445 590 735 880 1025 1170 1315 1460
4
300 588 877 1166 1455 1744 2033 2322 2611 2900
MIN.B/MIN.B'
No. of channels
4
5
6
7
8
50
60
70
80
90
100 110
86
105 124 143 162 181 200
, the Go/No Go
9
10
Unit : ms