Main display
Displays the measured data, setting data for various modes, or alarms, etc., for the
selected counter board.
Selected channel indicator
Choose one of the following with the initial setting.
1. Upper : A, Lower : B
2. Upper : A
3. Upper : A + B
• A
• In order to carry out calculations such as A–B or –A+B, change the direction of A or B to
Preset status indicator
When P is displayed the preset value is set.
Peak-hold indicator
When MAX/MIN/P-P is displayed the data shown is the maximum/minimum/maximum–
minimum value.
When neither of them is shown, the current value is displayed.
Comparator's upper limit setting indicator
Displayed when the comparator's value has been set. The upper digits are the upper limit.
Comparator's lower limit setting indicator
Displayed when the comparator's value has been set. The lower digits are the lower limit.
• Up to three different comparator's upper and lower limits settings (CPH1 to CPH3 and
Key lock indicator
"
Preset status indicator
Peak-hold indicator
P
A
P
B
Selected channel indicator
: data from measuring probe, input channel A (counter board)
B
: data from measuring probe, input channel B (counter board)
A+B : sum of data from channels A and B
"+" or "–" by the initial setting.
Note
Addition and subtraction can be performed only within the same counter board. These
operations cannot be performed across different counter boards.
CPL1 to CPL3) can be stored in the memory.
" is displayed when the keys are locked by the initial setting.
M A X M I N P ― P C P H1 2 3
mm
M A X M I N P ― P C P L1 2 3
mm
Key lock indicator
Comparator's upper
limit setting indicator
Comparator's lower
limit setting indicator