Accessories, transducers, and cables other than those specified, with the exception of transducers and
cables sold by the manufacturer of the medical electrical equipment as replacement parts for internal
components, may result in increased emissions or decreased immunity of the medical equipment and
model or part number.
The V.A.C.Via
Therapy Unit should not be used adjacent to or stacked with other equipment and
that if adjacent or stacked use is necessary, the V.A.C.Via
normal operation in the configuration in which it will be used.
Recommended separation distances between portable and mobile RF communications equipment and the V .A.C.Via™ Therapy
Unit
The V .A.C.Via
Therapy Unit is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer
or the user of theV .A.C.Via
can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the V .A.C.Via
the communications equipment.
Rated maximum
output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separate distance d in metres (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1, At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2, These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from surfaces, objects
and people.
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The V .A.C.Via
Therapy Unit is intended for use in an electromagnetic environment specified below. The customer or user of the V .A.C.Via
assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Conducted RF
IEC 61000-4-6
150K - 80 MHz
Radiated RF
IEC 61000-4-3
80 MHz - 2.5 GHz
NOTE 1, At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2, These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To asses the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the [ME EQUIPMENT OR ME
SYSTEM] is used exceeds the applicable RF compliance level above, the V .A.C.Via
abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the V .A.C.Via
b) Over the frequency range 150kHz to 80 MHz,, field strengths should be less than 3 V/m.
Therapy Unit as recommended below, according to the maximum output power of
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d=1.17 x sqrt (p)
0.1
0.4
1.2
3.7
11.7
Compliance level
level
3Vrms
3 V
3 v/m
3 V/m
18
Therapy Unit should be observed to verify
m
80 MHz to 800 MHz
d=1.17 x sqrt (p)
0.1
0.4
1.2
3.7
11.7
Electromagnetic Environment Guidance
Portable and mobile RF communications equipment should be
used no closer to any part of the V .A.C.Via
including cables, than the recommended separation distance
calculated from the equation application to the frequency of
the transmitter.
Recommended Separation Distance
d = (1.17) x sqrt(p)
d = (1.17) x sqrt(p) 80 MHz to 800 MHz
d = (2.34) x sqrt(p) 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer and is
the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey (see note a) should be less than the
compliance level in each frequency range (see note b).
Interference may occur in the vicinity of equipment marked with
the following symbol:
Therapy Unit should be observed to verify normal operation. If
800 MHz to 2.5 GHz
d=2.34 x sqrt (p)
0.2
0.7
2.3
7.4
23.4
should
Therapy Unit,
Therapy Unit.