Monitor cutaneous responses of patients who are incapable
of reacting, communicating and/or who cannot sense
temperature a minimum of every 15 minutes or according to
institutional protocol.
Adjust air temperature or discontinue therapy when the
therapeutic goal is reached, if elevated temperatures are
recorded or if there is an adverse cutaneous response in the
warmed area.
Do not place the warming unit on a soft uneven surface,
such as a bed, or a visibly wet surface as the air intake may
become blocked and cause the warming unit to overheat,
compromising the warming unit's performance.
Perform all temperature testing of the warming unit with a
3M Model 22110 temperature test unit.
!
CAUTION: To reduce the risk associated with
electromagnetic interference (EMI) due to portable and
mobile radio frequency (RF) communications equipment:
The 3M Bair Hugger temperature management system has
been tested to be resistant to both EMI and electrostatic
discharge (ESD).
Install and put into service the 3M Bair Hugger temperature
management system according to the electromagnetic
compatibility (EMC) information provided in the Guidance
and Manufacturer's Declaration.
Should interference occur, move away from the portable or
mobile RF communications equipment.

Notices

1.
The Bair Hugger warming unit meets medical electronic
interference requirements. If radio frequency interference with
other equipment should occur, connect the warming unit to a
different power source.
2.
To avoid warming unit damage:
Use proper Electrostatic Discharge (ESD) procedures when
performing maintenance.
Do not modify this equipment without authorization from
the manufacturer.
Do not immerse the warming unit, warming unit parts
or accessories in any liquid or subject them to any
sterilization process.
Do not use cleaning solutions with greater than 80%
alcohol or solvents, including acetone and thinner, to clean
the warming unit or hose. Solvents may damage the labels
and other plastic parts.
Proper Use and Maintenance
3M assumes no responsibility for the reliability, performance, or
safety of the warming unit if the following events occur:
Modifications or repairs are performed by
unqualified personnel.
The warming unit is used in a manner other than that
described in the Operator's or Service Manual.
The warming unit is installed in an environment
that does not meet the appropriate electrical and
grounding requirements.
The warming unit not maintained in accordance with the
procedures described in the Service Manual.
Read Before Servicing Equipment
All repair, calibration and servicing of the warming unit require the
skill of a qualified, medical equipment service technician who is
familiar with good practice for medical device repair. If service does
not require the manufacturer's attention, the Model 775 Service
Manual provides the technical information needed to service the
warming unit. Perform all repairs and maintenance in accordance
with the instructions in the Service Manual. For additional service
information please contact 3M technical support.
GB / 34-8718-8811-0
Overview and Operation
Carrying
handle
Figure A. Model 775 Warming Unit – front view
IV pole clamp
Hose connection
Power cord
connection
Figure B. Model 775 Warming Unit – rear view
Warming Unit Power‑On‑Reset
The Model 775 warming unit automatically performs the following
power-on-reset sequence after you connect the warming unit to a
properly grounded power source:
Performs all self-test functions.
Illuminates indicator lights and all pixels in the
alphanumeric display momentarily.
Displays the text "MD 700's" and the software revision
level in the alphanumeric display.
Sounds the alarm (three low-level clicks).
Enters the Standby mode with the high fan setting selected
by default.
Note: The low fan setting indicator light will not illuminate
during power up.
If the warming unit loses power for less than 1 second, the warming
unit software will return the warming unit to the operating mode
you selected prior to the power loss. However, if the warming unit
loses power for longer than 1 second, the warming unit software will
reset when you restore power. The warming unit will then enter the
Standby mode with the high fan setting selected by default.
Alphanumeric
display
Control Panel
Hose
5