CLASSIFICATION
The Model 775 warming unit is intended for use in the electromagnetic environment specified below. The customer or the user of the Model
775 warming unit should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
Flicker emissions
IEC 61000-3-3
The Model 775 warming unit is intended for use in the electromagnetic environment specified below. The customer or the user of the Model
775 warming unit should assure that it is used in such an environment.
Immunity Test
IEC 60601 test level
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
Electrical
±2 kV for power
fast transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/output
lines
Surge
±1 kV line to line
IEC 61000-4-5
±2 kV line(s) to earth
Voltage dips, short
<5% U
interruptions and
(>95% dip in U
voltage variations
for 0,5 cycle
on power supply
40% U
input lines
(60% dip in U
IEC 61000-4-11
for 5 cycles
70% U
(30% dip in U
for 25 cycles
<5% U
(>95% dip in U
for 5 sec
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
10
Guidance and manufacturer's declaration – electromagnetic emissions
Compliance
Group 1
The Model 775 warming unit uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
Class B
The Model 775 warming unit is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Class A
Complies
Guidance and manufacturer's declaration – electromagnetic immunity
Compliance level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/output
lines
±1 kV line to line
±2 kV line(s) to earth
<5% U
T
T
)
(>95% dip in U
T
for 0,5 cycle
40% U
T
T
)
(60% dip in U
T
for 5 cycles
70% U
T
T
)
(30% dip in U
T
for 25 cycles
<5% U
T
T
)
(>95% dip in U
T
for 5 sec
3 A/m
Electromagnetic environment ‑ guidance
Electromagnetic environment ‑ guidance
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30%.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
)
hospital environment. If the user of the Model 775 warming
T
unit requires continued operation during power mains
interruptions, it is recommended that the Model 775 warming
unit be powered from an uninterruptible power supply or
)
T
a battery.
)
T
)
T
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.
3M™Bair Hugger™ Model 775 Warming Unit – Operator's Manual