Safety System
THERMOSTAT
Independent electronic circuit; thermal cutoff shuts the heater OFF
to ensure hose end air remains below 56°C (53°C ± 3°C typical);
back-up over-temperature detection at warming unit hose inlet.
ALARM SYSTEM
Over-temperature (≤56°C, 53°C ± 3°C typical): red Over‑Temp
indicator light flashes, alarm sounds, heater and blower shut
down, operating indicator lights turn OFF, and control panel
becomes unresponsive.
FAULT
Amber Fault indicator light flashes, alarm sounds.
OVERCURRENT PROTECTION
Dual input fused lines.
Electrical Characteristics
HEATING ELEMENT
1400 W Resistive
LEAKAGE CURRENT
Meets UL 60601-1 and IEC 60601-1 requirements.
BLOWER MOTOR
Operating speed: 4,700 rpm (high fan setting)
4,100 rpm (low fan setting)
Airflow: up to 48 cfm or 23 L/s
POWER CONSUMPTION
Peak: 1550 W
Average: 800 W
GB / 34-8718-8811-0
POWER CORD
15 ft., SJT, 3 cond., 13 A
15 ft., SJT, 3 cond., 15 A
4.6 m, HAR, 3 cond., 10 A
DEVICE RATINGS
110-120 VAC, 50/60 Hz, 11.7 A, or
220-240 VAC, 50/60 Hz, 7.2 A, or
100 VAC, 50/60 Hz, 15 A
FUSES
TYPE
Fast acting ceramic fuses, 250 VAC
AMP RATING
12 A (110 - 120 VAC)
8 A (220 - 240 VAC)
15 A (100 VAC)
OPERATING SPEED
F (Fast Acting)
BREAKING CAPACITY
15 A, 12 A: 750 A @ 250 Vac
15 A, 12 A: 10,000 A @ 125 Vac
8 A: 200 A @ 250 Vac
8 A: 10,000 A @ 125 Vac
CERTIFICATIONS
IEC 60601-1; IEC/EN 60601-1-2; UL 60601-1; CAN/CSA-C22.2,
No.601.1; EN 55011; EN 80601-2-35