Ritter R400
Client (dental practice)
Device
Device type:
Dentale Behandlungseinheit
Manufacturer:
Ritter Concept GmbH
Type:
Serial No:
Netzspannung:
..........V
Safety inspection according: EN 62353 / VDE 0751
Visual inspection
Protective earth conductor not demaged (for insulation class 1 only)
Cover and mechanical parts not demaged
Insulating parts not demaged
Device connecting cables including plugs without defects
Labels exist or completed, nameplate correctly
Others
Measurement
Protective earth resistance
Insulating resistance
Test voltage
Device leakage current (Equivalent device leakage current)
Encl. leakage current (Housing leakage current)
Leakage current of application part (Equivalent patient leakage current)
Test voltage leakage current (nominal Voltage ±10%)
Functional testing
Function and safety testing without defects
The unit has safety deficiencies:
Short-circuit L-N
Body circuit L/N-SL
Test cycle:
Next test date:
Tester:
Place, Inspection date
© Ritter Concept GmbH
C20-0003-2001
Inspection protocol
safety inspection according DIN EN 62353
...... months
...................
Signature
for
Contractor
Used measurement device
Device type:
Manufacturer:
Type:
Insulation class:
Type of applied parts:
measured value
Safety inspection
Electrical Safety Tester
1
Typ BF
limit value
O.K. / not O.K.
<0,3 Ω
>2 MΩ
~500 V
<10 mA
<5 mA
<5 mA
min.: -10% U
Netz
max. +10% U
Netz
Anhang
appendix
91