Herunterladen Inhalt Inhalt Diese Seite drucken

Terumo Xcoating CAPIOX RX Serie Bedienungsanleitung Seite 26

Inhaltsverzeichnis
Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 42
SAFETY MARGIN
Fig. 11
Safety Margin (using RX25/RX15R40 Reservoir)
20
18
16
14
12
10
8
6
4
2
0
0
200*
500
Volume in Reservoir (mL)
Safety Margin (using RX15R30 Reservoir)
20
18
16
14
12
10
8
6
4
2
0
0
70*
500
Volume in Reservoir (mL)
PERFORMANCE DATA RX15
Fig. 12
O
Transfer Rate (in vitro)
2
400
Condition
Bovine Blood
Hb
: 12+1g/dL
300
Temp : 37+1°C
SvO
: 65%
2
PCO
: 45+5mmHg
2
200
B.E.
: 0+5mEq/L
100
0
0
1
2
3
Blood Flow Rate (L/min)
Fig. 14
Heat Exchanger Performance Factor (in vitro)
1.0
0.9
0.8
0.7
0.6
Condition
0.5
Bovine Blood
0.4
Hb
: 12+1g/dL
0.3
Tbi
: 30+1°C
Twi
: 40+1°C
0.2
Water flow rate : 15L/min
0.1
0.0
0
1
2
3
Blood Flow Rate (L/min)
PERFORMANCE DATA RX25
Fig. 16
O
Transfer (in vitro)
2
600
Condition
V/Q=1
Bovine Blood
Hb
= 12+1g/dL
500
SvO
= 65+5%
2
PvCO
= 45+5mmHg
2
B.E.
= 0+5mEq/L
400
Temp = 37+1°C
300
200
100
6
0
2
4
8
Blood Flow Rate (L/min)
10
1000
1500
2000
1000
1500
Fig. 13
400
V/Q=1.0
300
200
100
0
4
5
0
Fig. 15
150
100
50
0
4
5
Fig. 17
CO
Transfer (in vitro)
2
Condition
V/Q=2
600
Bovine Blood
Hb
= 12+1g/dL
SvO
= 65+5%
2
PvCO
= 45+5mmHg
500
2
B.E.
= 0+5mEq/L
V/Q=1
Temp = 37+1°C
400
300
V/Q=0.5
200
100
6
0
2
4
8
Blood Flow Rate (L/min)
* Minimum operating blood level in
CAPIOX RX25RE, RX25RW,
RX15RW40, RX15RE40 Hardshell
Reservoir.
The graph indicates the time in which the
blood volume in the reservoir
2 L/min
decreases to minimum operating blood
3 L/min
level of 200 mL at each flow rate. If the
blood volume in reservoir is below
4 L/min
200 mL, bubbles may pass to the
5 L/min
oxygenator module.
2500
* Minimum operating blood level in
CAPIOX RX15RW30, RX15RE30
Hardshell Reservoir.
The graph indicates the time in which the
blood volume in the reservoir decreases
to minimum operating blood level of 70 mL
at each flow rate. If the blood volume in
reservoir is below 70 mL, bubbles may
pass to the oxygenator.
1 L/min
2 L/min
3 L/min
4 L/min
2000
CO
Transfer Rate (in vitro)
2
Condition
Bovine Blood
V/Q=2.0
Hb
: 12+1g/dL
Temp
: 37+1°C
SvO
: 65+5%
2
V/Q=1.0
PCO
: 45+5mmHg
2
B.E.
: 0+5mEq/L
V/Q=0.5
1
2
3
4
5
Blood Flow Rate (L/min)
Blood Side Pressure Drop (in vitro)
20
Condition
Bovine Blood
Ht
: 35%
Temp : 37°C
10
0
0.5
1.5
3
5
Blood Flow Rate (L/min)
Heat Exchanger
Fig. 18
Performance Factor
(in vitro)
1.0
0.9
0.8
0.7
0.6
0.5
Condition
Bovine Blood
0.4
Hb = 12+1g/dL
Tbi = 30°C
0.3
Twi = 40°C
0.2
Water flow rate = 15L/min
0.1
6
0
2
4
Blood Flow Rate (L/min)
Fig. 19
Blood Side Pressure
Drop (in vitro)
200
150
Condition
Bovine Blood
Hb
= 12+1g/dL
100
B.E.
= 0+5mEq/L
Temp = 37°C
50
0
2
4
6
8
Blood Flow Rate (L/min)
20
10
8
Inhaltsverzeichnis