Full-body eligible – Post-MRI scan
Patient feedback
Return to original therapy
settings
2020-01-01 MRI guidelines for Medtronic deep brain stimulation systems 37601 37602 37603
Table 6. Full-body eligible – Post MRI scan
Verify that the patient has not experienced adverse
effects as a result of the MRI scan.
Contact Medtronic to report any adverse effects.
When the scan is complete, instruct the patient to
see the DBS physician to return therapy to the
original settings.
Or, if the patient has brought a patient programmer
to the MRI appointment, instruct the patient (outside
of the scanner room) to exit MRI mode, turn therapy
on, or return therapy to the original settings.
Note: If the programmer cannot synchronize with
the neurostimulator or cannot turn therapy back on
or displays a screen with the letters "POR" on it,
instruct the patient to follow up with the DBS
physician. Contact Medtronic to report the POR
event.