Frontal Handpiece

See illustration 005.
A. Handpiece spike and Irrigation Bag.
B. Pump.
b2. Pump marks:
C. Handpiece (stereo) plug.
D. Irrigation tubing clamp.
E. Frontal handpiece.

Setup and Operation

1. Mount the console on the irrigation pole.
Verify that the On/Off switch on the front of the console is set to
the Off (O) position.
2. Connect the power cord to the input on the rear of the console and
to a medical grade outlet.
3. Verify the system status indicator (green LED) is NOT illuminated.
Insert handpiece (stereo) plug into console (stereo) jack.
NOTE: If console power is ON (LED is illuminated) a momentary
activation of the peristaltic pump will occur when inserting the
handpiece (stereo) plug.

Irrigant

1. Hang irrigation solution bag(s) on the irrigation pole.
2. Adjust the tubing clamp(s) on the irrigation tubing to the closed
position.
For single irrigation bag/handpiece configuration, close tubing
clamp (A) at handpiece then spike bag. See illustration 006.

Pump Setup

1. Load irrigation pump tubing into the peristaltic pump. See
illustrations 007 (Pump marks) and 008 (Tube fittings).
(1) Lift the lid on the peristaltic pump and insert the tubing per
illustrations.
(2) Center the tubing on the pump head.
b1. Tube fittings:
Install tubing with one tube fitting on each side of pump.
Allow tubing to lay across roller naturally, do not pull or
stretch tubing.
b2. Pump marks:
Install tubing with one mark on each side of pump.
Allow tubing to lay across roller naturally, do not pull or
stretch tubing.
(3) Close the lid on the pump head.
2. Adjust the slide clamp on the irrigation tubing to the OPEN
position.
3. Toggle the On/Off switch on the front of the console to the On (l)
position and verify the system status indicator is illuminated.
4. Direct the tip of the handpiece over a sponge or collection tray.
Depress the actuation (trigger) button until irrigant is observed
flowing at the tip.
5. Release the actuation button and adjust the slide clamp on the
irrigation tubing to the closed position.

Operative Procedure

Maxillary Handpiece

Maxillary Handpiece Tip Articulation

See illustration 009
1. Rotating the tip articulation wheel to the right will turn the spray
nozzle to the right.
2. Centering the tip articulation wheel will center the spray nozzle.
3. Rotating the tip articulation wheel to the left will turn the spray
nozzle to the left.

Maxillary and Frontal Spray Nozzle Operation

Maxillary Handpiece and Frontal Handpiece
See illustration 010 (Maxillary) and 011 (Frontal)
The "Spray Pattern" of the Hydrodebrider® handpieces is from the center
of a diameter to the circumference of said diameter (size of diameter
increases with distances).
Note: There is a 180° difference in the 0°, 90°, 180°, and 270°
marks on the two handpieces.
1. Set starting position of spray pattern by rotating nozzle rotation
wheel until desired angle (in degrees) is visible in view window on
top of handpiece.
Xomed® Hydrodebrider® System
2. Depress activation button to activate pump and begin spray pattern.
3. Turn nozzle rotation wheel clockwise/counter-clockwise adjusting
spray pattern.
4. Release activation button to stop spray.

Post-Operative Procedure

1. Power down:
Turn the power switch to the Off (O) position.
2. Close all tubing clamps.
3. Disconnect/Remove:
Handpiece stereo plug from console.
Suction tubing from the suction source.
Remove the tubing from peristaltic pump head.
4. Discard the irrigation bag, handpiece and tubing, following health-
care facility guidelines on proper disposal of contaminated materials.
Do Not Reuse These Components.

Cleaning

Console

Wipe down the console and pump with standard disinfectant.
Do not use alcohol, other solvents, or abrasive cleaners.
Dry the product with a clean, non-abrasive cloth.
Allow to thoroughly dry before storing.

Service

If the Xomed® Hydrodebrider® System does not function properly, refer
to the "Troubleshooting" section of this manual. If troubleshooting
attempts do not correct the problem, contact Medtronic Xomed
Customer Care; or for international, contact your local Medtronic
Xomed office or distributor.
The Xomed® Hydrodebrider® System is not user-serviceable, and no
preventive maintenance or re-calibration procedures are required except
for the directions listed in this manual.

Technical Specifications

This device meets IEC/EN 60601-1-2: 2nd and 3rd edition requirements.
This device complies with IEC/EN 60601-1: 2nd and 3rd edition, ANSI/
AAMI ES 60601-1, CSA 22.2 No. 60601-1
Size
Console
Maxillary Handpiece
Frontal Handpiece
Weight
Console
Maxillary Handpiece
Frontal Handpiece
Recommended Installation
Electrical
Input Voltage
Internal Fuse
Medical Equipment
Classification
Mode of Operation
Protection Against Water
Ingress
Console
Irrigation Cycle Control
Duty Cycle
Front Console Control
Front Console Indicators
Authorized Service
Organization
17.8 cm x 26.7 cm x 24.1 cm
24.1 cm x 4.6 cm x 11.7 cm
22.86 cm x 2.9 cm x 8.4 cm
2.3 kg
80 g
33 g
Pole Mounted
100 - 240 VAC, 50/60 Hz, 40 VA Max
5X20, T, 1AL, 250 V
Class I Equipment Type BF Applied Part
Short-Time Operation
IPX1
Activation Button/trigger
Maximum On 5 Seconds
Minimum Off 5 Seconds
Power switch
Power/System Indicator