Herunterladen Inhalt Inhalt Diese Seite drucken

Switching From A Mymode Back To Basic Mode; Changing Prosthesis Settings; Using Remote Control To Change Prosthesis Settings - Ottobock C-Leg 3C98-3 Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für C-Leg 3C98-3:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Use

10.2.4 Switching from a MyMode back to basic mode

Information on switching
Regardless of the configuration of additional MyModes in the adjustment software, it is always possible to
switch back to basic mode (mode 1) with a motion pattern.
It is always possible to switch back to basic mode (mode 1) by connecting/disconnecting the battery charger.
Before the first step, always check whether the selected mode corresponds to the required motion type.
Requirements for successful switching using motion patterns
The following points must be observed to carry out switching successfully:
Stretch out the prosthetic leg to the back and bounce on the forefoot while maintaining contact with the floor.
Weight must be placed on the forefoot while bouncing.
The load may not be taken off the foot fully when relieving the load.
Switching process
1) While maintaining constant contact with the floor and with the prosthetic leg extended, bounce on the forefoot
at least 3 times but not more than 5 times.
→ A beep signal will sound to confirm that the motion pattern has been recognised.
2) Move the prosthetic leg in next to the contralateral leg, set it down and keep it still for approx. 1 second.
→ A confirmation signal will sound to indicate that the prosthesis has successfully switched over to basic mode.
INFORMATION: If this confirmation signal does not sound, the leg with the prosthesis was not cor­
rectly repositioned next to the contralateral leg or the requirements were not observed when bouncing
the foot. Repeat the process to correctly switch to the desired mode.

10.3 Changing prosthesis settings

Once an active connection to a prosthesis has been established, the settings of the MyMode currently active can
be changed using the remote control or app.
Information for changing the prosthesis settings
Before changing settings, always check the display of the remote control or the remote control app in the main
menu to make sure the correct prosthesis has been selected. Otherwise, parameters could be changed for the
wrong prosthesis.
It is not possible to change prosthesis settings nor to switch to a different mode while the prosthesis battery is
being charged. Only the status of the prosthesis can be called up. Instead of the
appears in the bottom row of the screen.
The prosthetist's setting is located in the middle of the scale. After making adjustments, you can restore this
setting by moving the slider control into the middle.
Prosthesis settings should be optimally set using the adjustment software. The remote control or app is not
intended for use by the prosthetist in order to set up the prosthesis. The day-to-day behaviour of the prosthesis
can be changed to a certain degree when the patient uses the remote control or app (e.g. as they become
accustomed to the prosthesis). The prosthetist can use the adjustment software to trace these changes at the
patient's next appointment.
If the settings of a MyMode are to be modified, one must first switch to this MyMode.

10.3.1 Using remote control to change prosthesis settings

33
1) Use the
,
keys to select the menu option "DEVICE SETTINGS" in the main menu and confirm this with
the  key.
→ A list appears with the parameters for the currently active mode (see fig. 34).
2) Use the
,
keys to select the parameter you want and confirm this with the
84 | Ottobock
34
35
 symbol, the
key.
C-Leg 3C98-3/3C88-3
 symbol

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

C-leg 3c88-3

Inhaltsverzeichnis