Herunterladen Inhalt Inhalt Diese Seite drucken

Warnings/Error Signals - Ottobock C-Leg 3C98-3 Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für C-Leg 3C98-3:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Beep signal
none
1 x short
Mode switching
Beep signal
Vibration sig­
nal
1 x short
1 x short
1 x short
1 x short
1 x short
1 x short
2 x short
2 x short
3 x short
3 x short

15.2.2 Warnings/error signals

Error during use
Beep signal
10 x 1 sec., then switch­
ing to empty battery mode
Beep signal
10 x short
30 x long
Error while charging the product
LED on
LED on
power sup­
battery
ply
charger
C-Leg 3C98-3/3C88-3
Vibration signal
3 x short
1 x before beep signal
Additional action performed
Mode switching with the remote
control
Bouncing on the forefoot
Prosthetic leg moved next to con­
tralateral leg, set down and kept
still for 1 second
Prosthetic leg moved next to con­
tralateral leg, set down and kept
still for 1 second
Prosthetic leg moved next to con­
tralateral leg, set down and kept
still for 1 second
Vibration signal
3 x long
5 x long
10 x long
Vibration signal
1 x short at interval of
approx. 5 seconds
10 x for 3 seconds each
Intermittent for 5 minutes
Error
Country-specific plug adapter not fully
engaged on power supply
Non-functional socket
Defective power supply
Event
Charging mode started (3 sec. after connection of battery
charger)
Battery charger disconnected after start of charging mode
Event
Mode switching with the remote control
carried out
Bouncing pattern recognised
Switch to basic mode (mode 1) carried
out.
Switch to MyMode 1 (mode 2) carried
out.
Switch to MyMode 2 (mode 3) carried
out.
Event
Battery charge level under 25%
Battery charge level under 10%
Battery charge level under 5%
Event
Overheated hydraulic unit Reduce activity
Battery
charge
level
under 5 %
Severe error
e.g. failure of valve drives
Possibly no switching into
safety mode.
Resolution
Check whether the country-specific plug
adapter is fully engaged on the power
supply.
Check socket with another electric appli­
ance.
The battery charger and power supply
must be inspected by an authorised
Ottobock Service Centre.
Appendices
Required action
Charge battery
Restricted walking possible.
Attempt to reset this error by
connecting/disconnecting the
battery charger.
If the error persists, it is no
longer permissible to use the
product. The product must be
inspected by an authorised
Ottobock Service Centre.
Ottobock | 91

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

C-leg 3c88-3

Inhaltsverzeichnis