Herunterladen Inhalt Inhalt Diese Seite drucken

Information On Alignment/Adjustment - Ottobock C-Leg 3C98-3 Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für C-Leg 3C98-3:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

NOTICE
Mechanical stress on the power supply/battery charger
Lack of proper charging functionality due to malfunction.
► Do not subject the power supply/battery charger to mechanical vibrations or impacts.
► Check the power supply/battery charger for visible damage before each use.
NOTICE
Operating the power supply unit/charger outside of the permissible temperature range
Lack of proper charging functionality due to malfunction.
► Only use the power supply unit/charger for charging within the allowable temperature range. The section
"Technical data" contains information on the allowable temperature range (see Page 94).
NOTICE
Independent changes or modifications carried out to the battery charger
Lack of proper charging functionality due to malfunction.
► Have any changes or modifications carried out only by Ottobock authorised, qualified personnel.

4.6 Information on Alignment/Adjustment

CAUTION
Use of unsuitable prosthesis components
Falling due to unexpected behaviour of the product or breakage of load-bearing components.
► Use the product only in combination with components listed in the section "Combination possibilities" (see
Page 52).
CAUTION
Improper assembly of the screw connections
Falling due to breakage or loosening of the screw connections.
► Clean the threads before every installation.
► Apply the specified installation torque values.
► Observe the instructions for securing the screw connections and the use of the correct length.
CAUTION
Incorrect alignment or assembly
Falling due to damage to the prosthesis components.
► Observe the alignment and assembly instructions.
CAUTION
Errors during prosthesis alignment
Falling due to breakage of load-bearing components.
► At maximum extension (reached under full load!), it is essential to maintain a minimum distance of 3 mm (1/8")
between the hydraulic unit and the socket.
► At maximum flexion and insofar as contact with the frame of the knee joint cannot be avoided (in case of volu­
minous residual limbs), the socket must lie flat against the frame. Soft cushioning on the socket will assist in
keeping the socket flat.
► If there is contact between the frame and the joint at maximum flexion, then the joint must be fitted with a flex­
ion stop (e.g. in the case of voluminous residual limbs).
C-Leg 3C98-3/3C88-3
Safety
Ottobock | 57

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

C-leg 3c88-3

Inhaltsverzeichnis