Herunterladen Inhalt Inhalt Diese Seite drucken

Walking Down A Ramp; Mymodes; Switching Mymodes Using The Remote Control Unit - Ottobock C-Leg 3C98-3 Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für C-Leg 3C98-3:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Use
4) Place the foot of the contralateral leg on the next step.
5) Place the foot of the prosthesis leg on the next step after that.

10.1.8 Walking down a ramp

Under increased flexion resistance, permit controlled flexion of the knee joint which lowers
the body's centre of gravity.
The swing phase is not triggered even though the knee joint is flexed.

10.2 MyModes

In addition to basic mode (mode 1), MyModes can be activated and configured with the adjustment software. They
can be called up by the patient using the remote control app, remote control or motion patterns.
These modes are intended for specific motion patterns or postures (e.g. inline skating, ...). Default settings for
these motion patterns and postures can be called up and individually adapted using the adjustment software.
Settings can also be adjusted by the patient using the remote control (see Page 86).

10.2.1 Switching MyModes using the remote control unit

INFORMATION
When the menu option "MODE" has the
MyMode settings cannot be retrieved.
INFORMATION
Bluetooth on the prosthesis must be turned on in order to use the remote control.
If Bluetooth is switched off, it can be turned on by turning the prosthesis upside-down or by connecting/discon­
necting the battery charger. Bluetooth is then turned on for approx. 2 minutes. The connection must be estab­
lished during this period. If required, Bluetooth on the prosthesis can be switched on permanently afterwards
(see Page 86).
Once a connection to a prosthesis has been established, the remote control can be used to switch between
modes.
30
1) Use the
,
 keys to select the menu option "MODE" in the main menu and confirm this with the  key.
2) Use the
,
 keys to select the mode you want and confirm this with the  key.
INFORMATION: If the list of modes was called up accidentally, the active mode can be selected again.
82 | Ottobock
 symbol next to it in the display of the remote control unit, it means that
31
32
C-Leg 3C98-3/3C88-3

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

C-leg 3c88-3

Inhaltsverzeichnis