Footswitch connection:
 Connect the footswitch into the receptacle on the front of the central unit ( Left – white)
 Place the footswitch on the ground.
While inserting the plug, align the guide on the connector in an upright position to fit into the corresponding female guide on the
E
™ (Fig. 13). Connectors have a spring retention system to help avoid accidentally disconnecting the unit's cords and cables.
NDODUAL
Switch on the device:
Electrical connections
Check the voltage present on your power line (100VAC to 240VAC).
Connect the power cord to E
of the unit.
Connect the power supply into the electrical outlet.
VII – SHUTTING DOWN THE DEVICE
Upon completion of the dental procedure:
- Switch the device off (O).
- Remove the instrument fastened to the contra-angle.
- Disconnect the contra-angle from the
- Disconnect the micro-motor cord from the
VIII – ROUTINE MAINTENANCE / STERILIZATION
Important:
The
E
micromotor and cord is not autoclavable.
NDODUAL

8.1 CONTRA-ANGLE MAINTENANCE

Refer to the manufacturer's instructions for your contra-angle.

8.2 ROTARY INSTRUMENT MAINTENANCE

Please refer to the manufacturer's instructions for your instruments (files).

8.3 DEVICE MAINTENANCE

Important:
Do not use an abrasive product to clean the device.
Do not allow cleaning/disinfecting spray to penetrate inside the device.
The device, micromotor and footswitch are not sterilizable.
Unplug the main unit from power source before cleaning or disinfecting. Do not use directly on liquids and sprays. Never immerse the
unit or its parts in liquid.
External surface of the console may be cleaned with a damp cloth or disinfectant wipe. The housing is not waterproof. Do not use
ultrasonic or steam sterilizers.
The
E
control unit and the footswitch must always be cleaned and disinfected after each procedure using alcohol, disinfectant
NDODUAL
agents or disinfectant wipes for dental practices.
The exterior of the footswitch may be cleaned by wiping with a soft cloth moistened with mild detergent or disinfecting solution.
Regular monitoring of the
E
NDODUAL
™ by inserting the female connector of the power supply into the male receptacle on the back
NDODUAL
E
micromotor.
NDODUAL
E
unit.
NDODUAL
unit is necessary in order to detect any problem.