IX – SHUTTING DOWN THE DEVICE
Upon completion of the dental procedure:
Remove
the irrigation line perforating pin from the bottle or bag of physiological saline.
Remove
the bottle or bag from the bracket.
Immerse
the irrigation line perforating pin in a vessel containing distilled water.
 Rinse the irrigation line and the contra-angle by running the flush function (flow 4) until the recipient and
irrigation lines are completely empty.
Disconnect
the irrigation line from the contra-angle.
Remove
the instrument fastened to the contra-angle.
Disconnect
the contra-angle from the micromotor.
Disconnect
the micromotor cord from the I-
Switch the device off (O).
X – ROUTINE MAINTENANCE / STERILIZATION
Important:
The "Implant" micromotor and cord are not autoclavable. The "Implant Ster" micromotor and cord
are autoclavable but not sterile upon delivery.
During sterilization, items made of different metals must not come into contact. Any contact would establish
galvanic couples and cause local damage. To avoid this happening, place the items in individual sterilizable
bags or in a sterilization box.
To maintain the sterility or asepsis of the accessories (contra-angle, "Implant Ster" micromotor and cord,
etc.), store them in sealed bags or containers suitable for medical practice.
The maintenance and/or sterilization instructions below must be applied before each use of the device.
Avoid using cleaning and disinfectant agents containing inflammable substances. If this is not possible, make
sure all such agent has evaporated and that there is no combustible matter on the device or its accessories
before it is switched on.
Sterilized items (handpiece holders, "Implant Ster" micromotor/cord, clips and autoclavable irrigation lines)
must be allowed to dry and cool to room temperature before they are reused.
Before sterilization, check that your autoclave is clean and that the water used is of adequate quality.
After each sterilization cycle, remove the items from the autoclave immediately in order to minimize the
risks of metal corrosion.