b. Control unit back panel
1. Main Power On/Off Switch. (Fig. 3 – 1)
2. Power Supply in. (Fig. 3 – 2)
c. Control footswitch
Activates the pre-selected maximum speed or start Forward/Reverse Movement (Fig. 4).
d. Technical characteristics
Console / micromotor / footswitch
Manufacturer
By Dental
Dimensions
170 x 72 x 130 mm
(W x H x D)
Materials
ASA
Micromotor
10000 rpm/Torque: 6.8 mNm
Noise
< 65 dBA
Input Voltage
9 VDC
Input Current
2.5 A
Power adaptor
Manufacturer
SL Power and AULT
Model
MENB 1030A0903C01
Input Voltage
100 – 240 VAC
Input Current
1A MAX
Not suitable for use in presence of flammable anesthetics or oxygen.
Conditions of use: Temp +18°C/+40°C (+64°F/+140°F) RH < 80%. Altitude restriction: 2,000 metres (6,500 feet).
Shipping and Storage conditions: Temp +5°C/+65°C (+41°F/+149°F) RH < 20-95% non condensing.
Life of the device: When used as prescribed = 5 years. No warranty claim may be made in the event that the failure happen soo ner or later
than indicated, because of the frequency of use, sterilization and maintenance.
IV– INSTALLATION / FIRST USE
4.1 UNPACKAGING THE DEVICE
Upon reception of the device, check for any damage caused in transit.
Contact your supplier if necessary.

4.2 RECOMMENDATIONS

Have your device connected to the power supply by a certified dental installation technician. The electrical connection of the
must comply with the standards applicable in your country.

4.3 INSTALLATION

Important:
Do not place the
Do not place the power cord and the footswitch cord in a wire cover or in a cable gland.
1. Remove the unit and its accessories from their original packaging and place on a flat surface. Do not install near heat sourc es, direct
or indirect.
2. Verify the contents of the box according to the section "3.1 PHYSICAL DESCRIPTION".
3. Do not kink the motor hose since it may damage it.
4. All accessories are supplied non-sterile.
5. The serial number on the unit's rear panel must be the same as the one on transport documents.
6. Allow at least 150mm free space around the unit for cooling ventilation.
E
close to or on top of another device.
NDODUAL
Protection
Class II
Type
B
Class MD
IIa rule 9 in Annex IX 93/42/EEC
Protective index
IP20
console
Protective index
IP65
micromotor
Protective index
IP20
footswitch
Output Voltage
9 VDC
Output Current
3000 mA
Class
II
E
NDODUAL