X - SHUTTING DOWN THE
DEVICE
Upon completion of the dental procedure:
- Remove the bottles or bags of physiological
saline or sterile water from the brackets.
- Remove the irrigation line perforating pins from
the bottles or bags.
- Immerse the irrigation line perforating pins in a
vessel containing distilled water.
- Rinse each irrigation line alternately, along with
the handpiece, by running the flush function
until the vessel and the irrigation lines are
completely empty.
- Remove the irrigation line clips.
- Disconnect the single-use irrigation lines from
the handpieces and dispose of them in a safety
container for soiled medical items.
- Disconnect the P
N
handpiece and the scaler cord.
EWTRON
- Switch the device OFF (O).
XI - ROUTINE MAINTENANCE/

STERILIZATION

Important:
The device must be switched OFF during
cleaning and disinfection procedures.
The only irrigation lines that can be sterilized
are those identified as sterilizable.
During sterilization, items made of different
metals must not come into contact. Any contact
would establish galvanic couples and cause
localized damage.
To avoid this happening, place the items in
individual sterilizable bags or in a sterilization
box.
16
2 handpiece or the
IEZOTOME
To maintain the sterility or asepsis of the
accessories (cords, handpieces, etc.), store
them in sealed bags or containers suitable for
medical practice.
The
maintenance
instructions below must be applied before each
use of the device.
Avoid using cleaning and disinfection agents
containing flammable substances.
If this is not possible, make sure that all the
agent has evaporated and that there is no
combustible matter on the device or its
accessories before it is switched on.
Sterilized items must be allowed to dry and
cool to room temperature before they are
reused.
Before sterilization, check that your autoclave
is clean and that the water used is of adequate
quality.
After each sterilization cycle, remove the
items from the autoclave immediately in order
to minimize the risk of metal corrosion.

11. 1 IRRIGATION LINE MAINTENANCE

Sterile irrigation line
The irrigation lines supplied by SATELEC are
disposable and must always be discarded after
use. Re-use of an irrigation line may result in
contamination of your patients and expose you to
de facto liability. Do not attempt to resterilize
the single-use irrigation lines. Do not attempt to
modify the irrigation lines.
Sterilizable irrigation line
Refer to the user's manual delivered with the
equipment.
and/or
sterilization