Fault detected: The motor doesn't work
Possible cause
Micromotor cable not connected properly
Contra-angle defective or not properly
calibrated
Wrong voltage
PCB damaged
Motor damaged
Connector pins and/or cable damaged
Fault detected: The motor stops suddenly
Possible cause
Micromotor /Cable damaged
Wrong voltage
PCB damaged
Fault detected: The Motor will not turn off
Possible cause
Broken connections inside
PCB damaged
Wrong voltage
X. ELECTROMAGNETIC COMPATIBILITY
Important:
The power cord, micromotor cord, and footswitch cord must be kept apart.
The
E
requires special precautions to be taken with regard to electromagnetic compatibility. It must be installed and prepared for
NDODUAL
use as described in chapter IV.
Certain types of mobile telecommunication devices such as mobile telephones are likely to interfere with the
separation distances in this paragraph must therefore be complied with.
The
E
must not be used near to or on top of another device. If this cannot be avoided, its operation under the conditions of use
NDODUAL
must be checked beforehand.
The use of accessories other than those specified or sold by the manufacturer as replacement parts may have the consequence of increasing
the emissions or decreasing the immunity of the
To avoid possible risks of electromagnetic interferences do not use other electro-medical instruments of similar nature near
E
™. The unit complies with the current normative electromagnetic radiation law.
NDODUAL
This device has been tested and found to comply with the emissions requirements of IEC 60601-1-2. These requirements provide
reasonable protection against harmful electromagnetic interference in a typical medical installation. However, high levels of
radio-frequency (RF) emissions from electrical devices, such as cellular phones, may disrupt the performance of this device. To
mitigate disruptive electromagnetic interference, position this device away from RF transmitters and other sources of
electromagnetic energy.
Risks from electromagnetic fields: The functions of implanted systems (such as pacemakers) can be influenced by
electromagnetic fields. Patients and users must inquire about implanted systems before using the motors and check the use. Do
not bring the motors close to the systems. Take suitable emergency precautions and immediately react to any changes in health.
Solution
Check to see that the micromotor cable is plugged into the connector on the front right
of the central unit, disconnect and reconnect.
Check the integrity of the contra-angle. Remove the contra-angle and perform
calibration.
Verify that the voltage matches the power supply. If it does, try to switch the central
unit off and then on again, or connect the device to another electrical outlet.
Immediately discontinue use of the device and contact SATELEC
Solution
Check to see that the micromotor cable is plugged into the connector on the front right
of the central unit, disconnect and reconnect.
Verify that the voltage matches the power supply. If it does, try to switch the central
unit off and then on again, or connect the device to another electrical outlet.
If you cannot find the cause of the malfunction, immediately discontinue use of the
device and contact SATELEC
Solution
Immediately discontinue use of the device and contact SATELEC
Verify that the voltage matches the power supply. If it does, try to switch the central
unit off and then on again, or connect the device to another electrical outlet.
E
™.
NDODUAL
®
after-sales-service.
®
after-sales-service.
®
after-sales-service.
E
™. The recommended
NDODUAL
16