Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: -These specifications may not be applicable in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and persons.
(a): The electromagnetic field strengths of fixed radiofrequency emitters, such as base stations for mobile telephones (cellular/cordless),
mobile radios, amateur radio, AM/FM radio broadcasts and TV broadcasts cannot be determined exactly by theory. To assess the
electromagnetic environment due to fixed radiofrequency emitters, an electromagnetic environment measurement must be made. If the
measured radiofrequency field strength in the immediate environment where the product is used exceeds the compliance level sp ecified
above, the performance of the product must be tested to verify whether it conforms to the specification. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the product.
(b): In the 150 kHz to 80 MHz frequency range, the electromagnetic field strengths must be less than 3 V/m.

10.4 RECOMMENDED SEPARATION DISTANCES

The device is intended for use in an electromagnetic environment in which radiated radiofrequency disturbances are controlled .
The device user and/or installer can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile radiofrequency communications equipment emitters and the device, according to the maximum output power of the equipmen t, as
recommended in the table below.
Rated max. power of the emitter (W)
0.01
0.1
1
10
100
For emitters rated at max. power not listed above, the recommended separation distance d in meteres (m) can be estimated usin g the
equation applicable to the frequency of the emitter, where P is the max. power rating of the emitter in watts (W) according the
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These specifications may not be applicable in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and persons.

10.5 CABLE LENGTHS

Cables and accessories
Maximum length
Handpiece cords
Footswitch cord
Less than 3m
Power supply cord
Separation distance in metres (m) according to emitter frequency
150 kHz to 80 MHz
P
d = 1.2
0.12 m
0.38 m
1.2 m
3.8 m
12 m
manufacturer.
RF emission, CISPR 1 - Class A/Group 1
Harmonic current emission: IEC61000-3-2
Voltage fluctuation: IEC61000-3-3
Immunity to electrostatic discharge: IEC61000-4-2
Immunity to electrical fast transients/bursts: IEC61000-4-4
Immunity to surges: IEC61000-4-5
Immunity to voltage dips, short interruptions and voltage variations: IEC61000-4-11
Immunity to conducted disturbances induced by radiofrequency fields: IEC61000-4-6
Immunity to radiated radiofrequency electromagnetic fields: IEC61000-4-3
80 MHz to 800 MHz
P
d = 1.2
0.12 m
0.38 m
1.2 m
3.8 m
12 m
Complies with:
800 MHz to 2,5 GHz
P
d = 2.3
0.23 m
0.73 m
2.3 m
7.3 m
23 m