Herunterladen Inhalt Inhalt Diese Seite drucken

Unpacking The Device; Recommendations; Installation; Commissioning - Acteon P5 Newtron Benutzerhandbuch

Vorschau ausblenden Andere Handbücher für P5 Newtron:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 52
IV - INSTALLATION

4. 1 UNPACKING THE DEVICE

On receipt of the device, check for any damage
caused in transit.
If necessary, contact your supplier.

4. 2 RECOMMENDATIONS

Have connections to the electrical and water
supply carried out by an approved dental
installation engineer.
The device should be connected to the electrical
supply in accordance with current standards.
The water supply system should satisfy the
quality criteria required for the practice of
dental treatment. In addition, it should be
equipped with a water supply cut-off system.

4. 3 INSTALLATION

Important:
Never rotate the handpiece connector on its
cord, otherwise you may damage the device.
Do not place the device close to or on top of
another device.
Do not place the power cord and the footswitch
cord in a wire cover or in a cable gland.
- Set up the control unit in the desired position.
- Check that the control potentiometer is set to
position O (off).
- Connect the device's water tube with its filter
to the water supply.
- Connect the mains plug to an electrical socket.
- Place the footswitch so as to make it easy to
operate with the foot.
- Check there is no trace of damp in the
handpiece connection, otherwise dry it (wipe it
and blow on it with the multifunction syringe).
- Connect according to model:
• the S
N
UPRASSON
EWTRON
to the cord connector, by lining up the
electrical contacts, taking care not to twist
the two parts;
• the S
N
UPRASSON
light handpiece to the cord connector, by
lining up the marks on the handpiece and the
cord connector, taking care not to twist the
two parts
Caution: do not connect a S
handpiece to the scaler cord of a P5 N
L
, nor a S
ED
UPRASSON
the scaler cord of a P5 N
- Place the handpiece on the stand.

4. 4 COMMISSIONING

Before using the device for the first time, we
recommend following the instructions below:
- After installation, operate the spray for a few
minutes in order to rinse the device's internal
water system: with the device operating at
minimum power (level 1) and with pressure
being applied to the footswitch.
- Disconnect the handpiece.
- Carry out maintenance and/or sterilization of
all equipment according to the procedures
defined in chapter 6.
(P5 N
) handpiece
EWTRON
L
(P5 N
EWTRON
ED
EWTRON
UPRASSON
N
L
handpiece to
EWTRON
ED
XS.
EWTRON
L
)
ED
N
EWTRON
EWTRON
9

Werbung

Inhaltsverzeichnis

Fehlerbehebung

loading

Diese Anleitung auch für:

P5 newtron led

Inhaltsverzeichnis