9.3 OPERATION FAULTS
Fault detected: The central unit does not switch on.
Possible cause
Power supply not connected properly
Unit has not been switched on
Wrong voltage
Broken connections inside
PCB faulty
Power interrupted
Fault detected: The central unit turns off and stops working suddenly
Possible cause
Power supply cable damaged
PCB damaged
Connections inside interrupted
Fault detected: The foot pedal doesn't start the motor
Possible cause
Micromotor cable not connected properly
Footswitch cable is not connected properly
Male connector of the micromotor damaged
Pins of the connector damaged
Wrong voltage
Fault detected: The display is not working properly
Possible cause
Contra-angle damaged
Wrong voltage
Possible interference
Broken connections inside
Display damaged
PCB damaged
Fault detected: The keyboard is not working properly
Possible cause
Keyboard damaged
Broken connections inside
Fault detected: The sound signals are not working properly
Possible cause
Broken connections inside
Solution
Make sure the power supply is properly plugged into the electrical outlet and the
connector on the back of the unit.
Check the back of the unit to see that the On button has been pressed.
Verify that the voltage matches the power supply. If it does, try to switch the central
unit off and then on again, or connect the device to another working electrical outlet. If
the unit does not turn on, verify the integrity of the fuse in the unit.
Immediately discontinue use of the device and contact SATELEC
Solution
After checking the correct connection to the electrical outlet and the absence of liquid
in the unit, try to power the central unit. If the device does not turn on, immediately
discontinue use of the device and contact SATELEC
Solution
Check to see that the micromotor cable is securely plugged into the connector on the
front right of the central unit, disconnect and reconnect.
Check that the footswitch cord is plugged into the connector on the front left of the
central unit, disconnect and reconnect.
Unplug the micromotor and make sure that the male connector is not damaged, wet or
damp.
Unplug the micromotor and the footswitch cable, make sure the metal pins on the male
connector are intact and not damaged and have not broken off and/or remained stuck in
the female connector mounted on the unit.
Verify that the voltage matches the power supply. If it does, try to switch the central
unit off and then on again, or connect the device to another electrical outlet.
Solution
Check the integrity of the contra-angle.
Verify that the voltage matches the power supply. If it does, try to switch the central
unit off and then on again, or connect the device to another electrical outlet.
Check to see if the central unit has been positioned too close to another electrical
device which can cause interference. In this case, move the interfering electrical device
further away from the central unit.
If you cannot find the cause of the malfunction, immediately discontinue use of the
device and contact SATELEC
Solution
®
Contact SATELEC
after-sales-service.
Solution
Immediately discontinue use of the device and contact SATELEC
®
after-sales-service.
®
after-sales-service.
®
after-sales-service.
®
after-sales-service.