C Beispiel für eine hygienische Reinigungsmethode
C.1 UNICORN-Methode für CIP/Sterilisierung/Spülung des Systems
96
4.50 InletA InletA10
5.00 InletA InletA1
5.00 End_Block
•
0.00 Block Fill_all_B_inlets
(Fill_all_B_inlets)
0.00 Base SameAsMain
0.00 Gradient 0.0 {%B} 0.00 {base}
0.00 InletA Closed
0.00 InletB InletB1
1.00 InletB InletB2
1.50 InletB InletB3
2.00 InletB InletB4
2.50 InletB InletB5
3.00 InletB InletB6
0.00 Gradient 3,50 {%B} 0.00 {base}
3.50 InletB Closed
3.50 InletA InletA1
3.50 End_Block
•
0.00 Block Fill_air_trap
(Fill_air_trap)
0.00 Base SameAsMain
0.00 AirTrap Bypass
1.00 AirTrap Out_through_drain
1.50 AirTrap Fill
2.00 AirTrap Inline
2.00 End_Block
•
0.00 Block Fill_filter_housing
(Fill_filter_housing)
0.00 Base SameAsMain
0.00 Filter Bypass
0.50 Filter Out_through_drain
1.00 Filter Inline
1.00 Message "Open manual filter valve" Screen "No sound"
1.00 Pause INFINITE {Minuten}
2.00 Message "Close manual filter valve" Screen "No sound"
2.00 Pause INFINITE {Minuten}
2.00 End_Block
•
0.00 Block Fill_all_column_valves
(Fill_all_column_valves)
0.00 Base SameAsMain
0.00 Column Bypass_Both
1.00 Column Bypass_Under
1.50 Column DownFlow
2.00 Column Bypass_Over
2.50 Column UpFlow
2.50 End_Block
•
0.00 Block Fill_all_outlets
(Fill_all_outlets)
0.00 Base SameAsMain
ÄKTAprocess Bedienungsanleitungen für das System 28-9353-86 Ausgabe AD