Handling, Functionality, Risks
8.5.2 Empty Battery Mode
CAUTION
Hazards when empty battery mode is activated
Depending on the settings (established by the CPO using C-Soft), the orthosis either selects "high flexion damp
ing" (high safety/reduced comfort) or "low damping" (less safety/higher activity) when empty battery mode is
activated.
In certain situations, this can lead to falling caused by:
► Altered damping behaviour of the orthosis.
→ The CPO has to inform the patient of the selected damping behaviour in empty battery mode (High or Low).
→ For the "low damping" setting in empty battery mode, the patient must have the necessary muscular and cog
nitive abilities to control a freely moving knee joint without stance phase stability.
→ From empty battery mode, operating mode (1st mode) can be reactivated by charging the orthosis.
The damping value activated in empty battery mode can be configured via C-Soft. Damping can be high (stance
phase damping value) or low (cycling mode).
8.6 Operating States/Error Signals
The orthosis joint module indicates operating states and error messages with beep signals.
8.6.1 Safety Mode Activation Indicator
Beep signal
1x short at intervals of approx. 1
second
4x short then pause, duration 30
seconds
1x short, interval 5 seconds, duration
10 minutes
Ongoing short beep signal
Continuous beep signal
After the problem has been corrected – insofar as this is possible – 1st mode can be activated by connecting and
disconnecting the battery charger.
74 | Ottobock
Error
Overheating
Critical device error
Critical device error
Highly critical device error
Correction
Cease normal use of the orthosis immedi
ately and check whether stance phase con
trol is active.
Do not resume using the orthosis until the
joint module has cooled down and the
orthosis no longer emits the warning.
Cease normal use of the orthosis immedi
ately and check whether stance phase con
trol is active.
The entire orthosis must be sent to the
authorised Ottobock Service Centre for
inspection or repair. Your CPO is the con
tact person.
Cease normal use of the orthosis immedi
ately and check whether stance phase con
trol is active.
This error signal can be deactivated by con
necting the battery charger.
If this error signal occurs again, the entire
orthosis must be sent to the authorised
Ottobock Service Centre for inspection or
repair. Your CPO is the contact person.
Cease normal use of the orthosis immedi
ately and check whether stance phase con
trol is active.
The entire orthosis must be sent to the
authorised Ottobock Service Centre for
inspection or repair. Your CPO is the con
tact person.
C-Brace