Built-in power supply Type L 30 S and
Type L 100 S (Figs. 1b and 1c)
Electronic stabilisation of output voltage
Automatic adaptation to the local mains voltage
of 90 – 250 V, continuous. Max. output 30 or
100 W. Power consumption 45 or 160 W.
Fuses
All versions of the stand are equipped with
safety fuses on both phases. The fuse module
(1.10) can be removed with a sharp object.
Safety data → back of microscope (1.10); code
no. → p. 63.
The fuse ratings apply for all mains voltages.
Fig. 2 Assembly tools
1 Crosstip screwdriver*, 2 3 mm hexagonal screwdriver,
3 2 mm centering keys*, condenser UCL/UCLP, 4 1.5 mm
centering keys*, 5 3 mm Allen key*, 6 2.5 mm Allen key*
(short version)
10
12 V.
On no account should you use fuses of dif-
ferent ratings.
External power units
Assembly of lamphousings, lamp change
→ p. 21; 23.
Gas discharge lamps (Hg, Xe → p. 24) are
always fed by external power units. For incident
light, the 100 W halogen lamp can alternatively
also be run on an external power unit, e.g. for
simultaneous transmitted and incident light
illumination. A separate instruction manual is
supplied for external power units.