Herunterladen Inhalt Inhalt Diese Seite drucken

Stryker InTouch Bedienungsanleitung Seite 35

Intensivpflegebett
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 104
MAIN MENU: SCALE (CONTINUED)
Delayed Weight Button
When the Delayed Weight button is pressed on the
Scale screen, Figure 5 . 1 will be displayed, at which time
a 15 second timer starts .
Note
The delayed time is set 15 seconds .
To calculate patient weight with a delayed start:
1 .
Press Delayed Weight button .
2 .
The 15 second timer starts .
3 .
The operator has 15 seconds to lift up any medical
devices or equipment that should not be calculated
in the patient's weight . This includes anything lying
on the bed surface or resting on the patient (i .e .,
hoses) .
4 .
After 10 seconds expires, an audible beep will
sound .
5 .
For the next 5 seconds, the bed will calculate the
weight of the patient lying on the bed . The operator
must remain lifting equipment described in step 3 .
6 .
After 5 seconds expires, another audible beep
will sound and the Delayed Weight screen will be
displayed as shown in Figure 5 .2 . The bed has
completed its calculation and the equipment being
lifted can be repositioned on the bed surface .
Lbs/Kg
The user can change the delayed weight measuring
value to Kg or Lbs by pressing on the "Lb/Kg" button
located at the bottom of the Delayed Weight screen as
shown in the Scale menu .
www.stryker.com
Footboard Operation Guide
2131-209-005 REV B
Message
Start weight
with delay
Figure 5.1
Weight History
118.4 lbs
JUL. 22, 2008
09:22 AM
Previous
Figure 5.2
Cancel
Next
Cancel
Return To Table of Contents
1-31

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis