BED POSITIONING DESCRIPTIONS

Note
The addition of accessories affects the motion of the bed .
Flat (0˚)
To activate the Flat Bed position, press on the Flat Bed Position button located on the Nurse Control .
the Fowler, Foot and Gatch sections in the flat position .
Trendelenburg or Reverse Trendelenburg, the bed will be set into motion such that either the head Bed Lift and the foot
Bed Lift will move in order to achieve the Flat Bed position .
Note
It is possible to move the Gatch, Head and Foot sections at the same time .
Standard Cardiac Chair
To activate the Cardiac Chair position, press on the Chair button located on the Nurse Control . Keep pressing the
control until the complete Cardiac Chair position is obtained . The Fowler and Gatch will simultaneously raise to full up
position and then the bed will position itself into Reverse Trendelenburg .
Enhanced Cardiac Chair
To activate the Enhanced Cardiac Chair position, press on the Chair button located on the Nurse Control . Keep pressing
the control until the complete Cardiac Chair position is obtained . The Fowler and Knee Gatch will simultaneously raise
to full up position and then the bed will position itself into Reverse Trendelenburg .
Note
There will be a 1 second stop between each chair position .
Trendelenburg
The Trendelenburg position places the bed Fowler to the low position while the foot section is set to the high position .
This position can be achieved by pressing the Trendelenburg button on the Nurse Control until the desired angle is
obtained . The bed height can be lowered while it is in Trendelenburg without changing the angle .
Reverse Trendelenburg
The Reverse Trendelenburg position places the bed Fowler to a high position and the foot section to a low position .
This position is achieved by pressing the Reverse Trendelenburg button on the Nurse Control Panel .
Note
• The maximum angle of inclination that can be achieved during the Trendelenburg and Reverse Trendelenburg
position is 15 degrees .
• Upon detection of the I . V . pole, the foot section of the bed can no longer be set into motion .
• A minimum clearance of 20 inches is required to allow the Trendelenburg and Reverse Trendelenburg positions
to be executed .
• These positions cannot be executed when one or more locks is activated .
Return To Table of Contents
1-22
Siderail Operation Guide
In the event that the Flat button is pressed when the bed is in
2131-209-005 REV B
This will place
www.stryker.com