APPENDIX B: TROUBLESHOOTING

Problem
Staining, discoloration
The shaver blade does not turn.
The handpiece will not run.
The Activate function is not operating smoothly.
The shaver blade/RF probe is not correctly
detected by the handpiece.
DCO-1394
Effective Date: 12-JAN-2022
Possible Cause
- Inadequate cleaning (for example, remaining protein residues)
- Cleaning solutions are contaminated or being reused
The shaver blade is not inserted properly.
The handpiece is not properly connected to the controller.
The cord is malfunctioning.
- The cable is malfunctioning.
- The Controller is malfunctioning.
The shaver blade/RF probe is not inserted properly.
The handpiece is malfunctioning.
Recommended Action
- Clean in accordance with section 10.1
- Always use freshly prepared solutions
- Repeat cleaning in accordance with section 10.1
Remove the blade and reinsert it according to the RELIGN Tricera System instructions for directions on
use of the Tricera Handpiece as part of the RELIGN Tricera System.
Remove and reconnect the handpiece cable to the controller according to the RELIGN Tricera System
instructions for directions on use of the Tricera Handpiece as part of the RELIGN Tricera System.
Send the handpiece in to RELIGN for repair.
- Remove and reconnect the handpiece to a different Controller. If the problem persists, send the Tricera
Handpiece to RELIGN for repair.
Remove the shaver blade/RF probe and reinsert it according to the RELIGN Tricera System instructions
for directions on use of the Tricera Handpiece as part of the RELIGN Tricera System.
Remove and reconnect the shaver blade/RF probe to another Tricera Handpiece to confirm the possible
cause. Send the malfunctioning Tricera Handpiece to RELIGN for repair.
LBL-6059 Rev D
Page 8 of 116