a Indications for Use
The Ranger blood/fluid warmer is intended to warm blood, blood products,
and liquids.
Patient Population and Settings
Adult and pediatric patients being treated in operating rooms, emergency trauma
settings, or other areas where blood/fluids are infused.

Explanation of Signal Word Consequences

WARNING: Indicates a hazardous situation which, if not avoided, could result in
death or serious injury.
CAUTION: Indicates a hazardous situation which, if not avoided, could result in
minor or moderate injury.
WARNING: To reduce the risks associated with air embolism:
• Never infuse fluids if air bubbles are present in the fluid line.
• Ensure all luer connections are tightened.
• Bubble trap does not eliminate air automatically. See instructions on "Removing
Air from the Disposable Set".
• Do not mount the Ranger blood fluid warming system higher than patient level,
as air embolism may result.
• Always keep bubble trap in a vertical position during infusion
WARNING: To reduce the risks of biohazard exposure, cross-contamination
or infection:
• Do not reuse. This product is designed for single patient use only.
• Do not use if caps are not in place.
• Ensure that all luer connections are securely tightened prior to priming set.
WARNING: To reduce the risk of leaking and loss of therapy:
• Do not reuse. This product is designed for single patient use only.
• Ensure that all luer connections are securely tightened prior to priming set.
WARNING: To reduce the risks associated with potential blood loss:
• Do not use in combination with an extracorporeal circuit.
WARNING: To reduce the risk associated with patient exposure to leakage current
that exceeds Type BF equipment limits:
• Do not use in combination with other medical equipment without verifying that
the total leakage current from the combined equipment does not exceed the
safety limits for Type BF equipment.
CAUTION: To reduce the risk of leaking and loss of therapy:
• Do not use with pressure, use only gravity.
• Do not remove warming cassette from warming unit while in use
CAUTION: To reduce the risks associated with air embolism:
• Do not inject or administer any fluid into air aspiration port.
CAUTION: To reduce the risks associated with warming platelets, cryoprecipitate,
or granulocyte suspensions:
• Follow the AABB Guidelines for the use of blood warming devices which
caution against warming when administering platelets, cryoprecipitate, or
granulocyte suspensions.
CAUTION: To reduce the risks associated with potential material toxicity:
• Do not use in combination with an extracorporeal circuit as the Ranger
blood/fluid warming system is not designed for use with a circulating blood
system; therefore, biocompatibility testing has not been completed to
assess genotoxicity.
A
B
C
G
Instructions for Use
Do not use if package has been previously opened or damaged. Refer to the 3M™
Ranger™ blood/fluid warming unit model 245 Operator's Manual for set up and
use of the warming system.
A
B
Figure 1: Illustration of Disposable Set
A. Inlet line
B. Blue Inlet pinch clamp
C. Fluid Warming Cassette
D. Needleless Injection port
E. Bubble trap with air aspiration port
F. White outlet pinch clamp
G. Patient Connection
Figure 2
Peel To Open
Instructions
Set-up and Priming the disposable set
1.
Adjust warming unit height to patient level or lower.
2. Open warming set pouch where indicated by the arrows (Fig. 2)
3. Before priming, slide the fluid warming cassette (Fig. 1, C) into the slot in the
warming unit.
4. Connect the inlet line (Fig 1, A) on the Ranger warming set to the fluid source.
- 2 -
F
D
E
C
Peel To Open
G
F
D
E