EN 61000-3-2
Power Line Harmonics
EN-61000-3-3
Power Line Flicker
7.3.2.

Immunity Test Summary

Specification
Minimum Test Level Required per EN
60601-1-2, 4
EN 61000-4-2
Air discharge up to ± 15 kV;
- Electrostatic
Contact discharge up to ± 8 kV
Discharge Immunity
EN 61000-4-3
Radiation field strength of 3V/m from 80 –
- RF Radiated Fields
2700 MHz
Immunity
(80% AM, 1 kHz)
EN 61000-4-4
Power line pulses of ± 2kV direct;
-Electrical Fast
I/O line pulses of ± 1kV
Transient Immunity
EN 61000-4-5
Power line surges of ± 2kV common, ± 1kV
-Lightning Surge
differential mode
Immunity
EN 61000-4-6
150 kHz - 80 MHz at 3 Vrms
-RF Common Mode
1 kHz 80% amplitude modulated
Immunity
EN 61000-4-8
Inductive loop at 50 Hz or 60Hz, to 30 amps
-Power Frequency
(rms) per meter
Magnetic Field
Immunity
EN 61000-4-11
Voltage Dips of
- Voltage Dips and
100% for 0.5 cycles (0, 45, 90, 135, 180, 225,
Short Interruptions
270 and 315)
100% for 1 cycle
30% for 25 cycles
10% for 250 cycles (5s)
7.3.3.

Recommended Separation Distances

The device is intended for use in the electromagnetic environment in which radiated disturbances are controlled. The customer or user of the
device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications
Equipment and the device as recommended below, according to the maximum output power of the communications equipment.
Below is the table of recommended separation distances between portable and mobile RF communications equipment and the device and systems
that are not life-supporting.
Max Output
Power
(watts)
0.01
0.1
1
10
100
7.3.4.
Guidance and Manufacturer's Declaration – Immunity – For Non-Life-Supporting Equipment
The Laser System is intended for use in the electromagnetic environment specified below. The customer or user of the device should ensure that
it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Conducted RF
3 Vrms
IEC 61000- 4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000- 4-3
80 MHz to 2.5 GHz
RSR-000886-000 (5)
Up to the 40
Harmonic
th
Less than or equal to 4% Maximum Relative Voltage Change;
Value of d(t) less than or equal to 3.3% for more than 500 ms
Edition
th
Separation (m)
150 kHz to 80 MHz
D=(3.5/V1)(Sqrt P)
0.1166
0.3689
1.1666
3.6893
11.666
Compliance
Level
(V1)Vrms
(E1)V/m
Test Level Completed
Air discharge up to ± 15 kV;
Contact discharge up to ± 8 kV
Radiation field strength of 3V/m from 80
– 2700 MHz
(80% AM, 1 kHz)
Power line pulses of ± 2kV direct;
I/O line pulses of ± 1kV
Power line surges of ± 2kV common, ±
1kV differential mode
150 kHz - 80 MHz at 3 Vrms
1kHz 80% amplitude modulated
Inductive loop at 50 Hz, to 30 amps
(rms) per meter
Voltage Dips of
100% for 0.5 cycles (0, 45, 90, 135,
180, 225, 270 and 315)
100% for 1 cycle
30% for 25 cycles
10% for 250 cycles (5s)
Separation (m)
80 to 800 MHz
D=(3.5/E1)(Sqrt P)
0.1166
0.3689
1.1666
3.6893
11.666
Electromagnetic Environment – Guidance
Portable and mobile communications equipment should be
separated from the laser by no less than the distances
calculated/listed below:
D=(3.5/V1)(Sqrt P)
D=(3.5/E1)(Sqrt P) 80 to 800 MHz
D=(7/E1)(Sqrt P) 800 MHz to 2.5 GHz
where P is the max power in watts and D is the
recommended separation distance in meters.
Field strengths from fixed transmitters, as determined by
an electromagnetic site survey, should be less than the
compliance levels (V1 and E1).
Interference may occur in the vicinity of equipment
containing a transmitter.
PASS
PASS
Compliance Status
PASS
PASS
PASS
PASS
PASS
PASS
PASS
Separation (m)
800 MHz to 2.5 GHz
D=(7/E1)(Sqrt P)
0.2333
0.7378
2.3333
7.3786
23.333
Page 16 of 20