Fiber Management

The side of the FXi console has an easy-to-use Fiber Manager Spool and Hand Piece Clip that allows the user to store the beam delivery
assembly while not in use. When therapy is discontinued or between treatment cycles, the user may rest the hand piece in the Hand Piece Clip.
For long term storage and during transport, the user may also wrap the fiber around the Fiber Manager Spool to avoid damage to the fiber.
Hand
Piece
Clip
Fiber
Slot
When wrapping the hand piece fiber around the fiber manager spool, the first fiber wrap should be fed into the Fiber Slot so the plastic Bifurcator,
which couples the finger switch and fiber optic cables, snaps into the slot. Five more wraps (for a total of six) should be made tightly around the
spool to provide enough length for the hand piece to feed into the Hand Piece Clip (Figure 20).
3.7.

Laser Safety Eyewear

Warning:
ALWAYS wear the protective eyewear supplied with this laser system when operating the device.
All personnel who are in the laser treatment controlled area MUST wear laser safety eyewear.
DO NOT remove protective eyewear until the operator returns the laser device to Standby mode.
If you suspect injury, such as direct exposure to the eyes, seek medical attention immediately.
Additional protective eyewear can be purchased from company webstore.
3.8.

Medical Grade Power Cord

Caution: Do not use any power cord other than the one provided with the system.
If a replacement is needed, additional power cords can be purchased from the company webstore.

4. Operation

4.1.

Safety Precautions

Anyone in the laser treatment controlled area during laser operation (including the operator, all assistants, and the patient) MUST wear the
protective eyewear supplied with this laser system.
Recommendation: DO NOT allow any nonessential personnel into the laser treatment controlled area during any laser procedure.
Recommendation: NEVER allow untrained personnel to operate this laser system unless directly supervised by a properly trained and
experienced individual.
RSR-000886-000 (5)
Nominal Ocular Hazard Distance (NOHD) Table
Image
Snap
in
Figure 20
Divergence (radians)
0.815
0.815
0.815
0.226
0.057
6
th
&
final
NOHD (meters)
0.6
0.4
0.4
1.1
4.9
Page 8 of 20