Herunterladen Inhalt Inhalt Diese Seite drucken

Treatment Technique; Maintenance; Cleaning; Cleaning The Console - Chattanooga LIGHTFORCE FXi LTC-1510-L-B6 Gebrauchsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 105
4.4.4.

Treatment Technique

Warning:
DO NOT allow any reflective objects to fall into or obstruct the path of the laser energy produced by this device. The operator, all
assistants, and the patient must remove all reflective objects (such as rings, metal watchbands, and jewelry) prior to treatment with
this device. Direct or indirect eye contact with scattered laser light from any reflective surface can cause serious, irreparable corneal
and/or retinal damage, and possible blindness to one or both eyes.
DO NOT pretreat tissue (e.g. with a heating pad or wet compress) before laser treatment. Tissue to be treated must be dry and at
normal temperature before laser treatment begins.
Use carefully. May cause serious burns. Do not use over sensitive skin area or in the presence of poor circulation.
During treatment, it is important to get feedback from the patient concerning their comfort level. The laser provides a soothing
warmth. Be cautious in cases where the patient has limited sensation or is not able to respond to increases in temperature and
consider treating at a lower power setting. If the patient expresses any discomfort, the power output should be reduced.
Treatment should be delivered to exposed skin. Do not treat through clothing or bandages. The treatment area
should be clean and free of surface dirt or oils.
Isopropyl alcohol solution (70% alcohol) may be used to clean all instrument surfaces that contact the patient (See
Section 5.1 for cleaning instructions). Be sure to allow the solution to evaporate before initiating laser emission.
For optimal delivery of the dose, keep the treatment head perpendicular to the skin surface. Continuously move
the treatment head at a rate of approximately 3-10 cm/sec (1-3 in/sec).
To ensure an even distribution of the dose to the treatment area, continuously move the treatment head over the
area first moving continuously back-and-forth horizontally, and then moving back-and-forth vertically in a grid
pattern (see Figure 25). Make sure to cover the entire area to be treated with this "painting motion".
Treatments can be applied in conjunction with stretching or gentle range of motion exercises.

Maintenance

5.
Warning: DO NOT attempt to gain access to any internal component. Doing so may cause serious and/or irreversible injury. THERE
ARE NO USER-SERVICEABLE COMPONENTS inside this laser device.
5.1.

Cleaning

Warning:
It is recommended to always turn off the laser system and unplug the power cord from the wall outlet before cleaning the system.
Always use protective eyewear and gloves when cleaning and disinfecting any equipment.
Clean the contact treatment heads (large massage ball and small massage ball) between treatments to prevent cross-
contamination and infections.
DO NOT use a treatment head immediately after cleaning with alcohol solution. Wait until the solution used in the cleaning
procedure completely evaporates. Not allowing it to dry may cause the laser to ignite alcohol solutions or vapors.
Caution: DO NOT clean the lens inside the hand piece. Doing so may damage the lens during laser emission.
5.1.1.

Cleaning the Console

Note: The laser system uses solid-state laser technology. It is important to keep the unit and accessories free from dust.
Step 1.
Before cleaning the console, turn off the laser system and unplug the power cord from the wall outlet.
Step 2.
The exterior surfaces on the unit and treatment heads can be cleaned by wiping with a soft, non-fibrous wipe (e.g., Kimwipes
moistened with isopropyl alcohol solution.
Ensure the solution is only 70% alcohol. Solutions of more than 70% alcohol can cause product damage.
Avoid using soiled or gritty cloths.
Step 3.
After cleaning, ensure that all cleansers have been removed and the parts are dry before use.
5.1.2.

Cleaning the Large Massage Ball

Step 1.
Remove the elastic band from the large massage ball treatment head.
Step 2.
Unscrew the top portion of the large massage ball treatment head (Figure 26).
Step 3.
The ball can now be removed for cleaning. Use a soft, non-fibrous wipe (e.g., Kimwipes
70% isopropyl alcohol solution. Do not wipe with a dry cloth, always dampen the cloth before wiping.
Recommendation: Be careful when unscrewing the top portion of the treatment head. It is recommended that this be done over a table to
reduce the chance of dropping the ball directly on the floor.
Step 4.
Ensure that all cleansers have evaporated and the parts are dry before re-assembly and use.
Step 5.
Reassemble treatment head by replacing the massage ball and screwing the top portion securely on the treatment head.
Caution: Dropping the ball, may damage the ball and plastic housing. If significant damage has occurred, discontinue use. Ensure the ball
is clean and free of debris that could cause damage over continued use.
RSR-000886-000 (5)
Figure 25
®
) moistened with
Figure 26
Page 12 of 20
)
®

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis