To use Walk Away mode – connect the AC power adapter to the Analyzer and a power source
Step 6B: Starting Walk Away mode
•
Turn the Analyzer on by pressing the blue power button once.
•
When the display window reads:
"INSERT TEST DEVICE OR DOUBLE-CLICK FOR WALK AWAY MODE,"
Double-click the blue power button.
–
Step 7B: Add the specimen to the test device
•
When the display window reads "ADD SPECIMEN TO TEST DEVICE AND
INSERT IMMEDIATELY":
–
Invert the tube, holding it vertically (approximately one inch above the BD Veritor
System Strep A device sample well).
–
Gently squeeze the ridged portion of the tube, allowing three (3) drops of the
processed specimen to dispense into the sample well of a labeled BD Veritor
System Strep A device.
NOTE: Squeezing the tube close to the tip may cause leakage.
Step 8B: Start the development and reading sequence
•
Immediately insert the test device into the slot on the right side of the Analyzer.
The test device must remain horizontal to prevent spilling the specimen out of
the sample well.
•
"DO NOT DISTURB TEST IN PROGRESS" appears in the display window. Automatic
timing of the assay development, image processing and result analysis begins.
•
A countdown timer in the display window shows the remaining analysis time.
•
Do not touch the Analyzer or remove the test device during this process. Doing
so will abort the assay analysis.
Step 9B: Record the Result
• When analysis is complete, the test result appears in the display window.
ATTENTION: TEST Results are NOT maintained in the display window when the device
Using a BD Veritor Plus Analyzer in "Walk Away" mode:
(with no optional module installed)
is removed or if the Analyzer is left unattended for more than
60 minutes (if the AC power adapter is connected).
7