Step 6A: Adding the specimen
•
Invert the tube and hold the tube vertically (approximately one inch above the labeled
BD Veritor System Strep A device sample well).
•
Gently squeeze the ridged body of the tube, dispensing three (3) drops of the processed
specimen into the sample well of a labeled BD Veritor System Strep A device.
NOTE: Squeezing the tube too close to the tip may cause leakage.
Step 7A: - Timing development
After adding the sample, allow the test to run for 5 minutes before inserting into the
•
BD Veritor Instrument.
•
NOTE: If running test under laminar flow hood or in an area with heavy ventilation, cover test
device to avoid inconsistent flow.
Step 8A: Using the BD Veritor Instrument
•
During incubation time, turn the BD Veritor Instrument on by pressing the power
button once.
•
Insert assay device when 5 minute assay development time is complete.
•
Follow the on-screen prompts to complete the procedure. The status of the assay
analysis process appears in the display window.
Step 9A: Record the Result
• When analysis is complete, the test result appears in the display window.
ATTENTION: TEST Results are NOT maintained in the display window when the device
Using a BD Veritor Reader or Analyzer in "Analyze Now" mode:
is removed or if the Analyzer is left unattended for more than
60 minutes (if the AC power adapter is connected).