Patient Population
The ultrasonic and electric motor functions can be used with the following patients without consideration of physical characteristics such a s
weight (except for children), age, size, sex and nationality:
- Children
- Adults
The ultrasonic and electric motor functions are prohibited to the following population:
- Infant
- Pregnant and lactating women
- Patients with medical complications
- Allergic patients
- Patients located on a clinical site which does not enable to carry out the treatment.
The patient must be calm, relaxed, motionless, ideally lying on a dental chair.
Body parts or tissue types covered
The clinical use of the ultrasonic and motor functions has to be limited to the buccal cavity of the patient.
Principle of operation of Device
Ultrasonic: An electrical signal issued by the ultrasonic generator is supplied to a ultrasonic dental handpiece which is connected to the
module located inside the dental unit through a cord.
The handpiece includes a piezoelectric ceramic transducer that transforms the electrical signal into ultrasonic vibrations.
The mechanical vibrations are transmitted to a dental instrument screwed onto the end of the transducer.
Motor: A rotating motor coupled with a dental contra-angle enables to drive dental rotating and reciprocating files.
Significant performance characteristics
Ultrasonic: Ultrasonic vibration of the instrument fixed at the end of the dental handpiece:
- Vibration frequency ≥ 28 kHz
- Amplitude of motion: ≤ 200 µm
Motor: Rotation speed of the instrument fixed at the end of the contra-angle:
- Maximum speed 10.000 rpm
- Maximum torque 10 Ncm
Applied parts
Dental ultrasonic handpiece
Dental ultrasonic instruments, tips
Dental contra-angle handpiece
Endodontic files
Use
The medical device is not restricted to a number of uses.
Depending on the number of repeated use, the pressure and the selected power, the tips and files may exceptionally break. It is
recommended to use a succion system during use of the medical device.
4. INSTALLATION
Carefully remove the unit and the accessories from their packaging and place them properly on a flat surface in the desired location.
Before set-up, check that all parts are present in the box:
 Central unit
 Footswitch (Fig. 4)
 Wallplug power supply
 Ultrasonic handpiece
 Sterijoint
 Endo MicroMotor / Cord
 Irrigation clips
 Irrigation bracket
 Irrigation line
 Handpiece holder
 Tip wrench
 Set of tips
 User manual
- Teengers
- Seniors
8