5. STEP-BY-STEP INSTRUCTIONS

Notes
 Before and after each use, your device must be disinfected with appropriate products.
 Before each procedure, it is essential to make sure that the accessories to be used have been cleaned, disinfected and sterilized.
Before turning on the unit, check if you have completed the installation of all components.

5.1. FIRST USE

When first switched on, the current software version appears on the screen for few seconds:
ACTEON
ENDOCENTER
V.1.0
And the values set by manufacturer are displayed:
ULTRASONIC
Power range:
Setting:
According to the expected use, press the "MODE" key to toggle from "Rotating" Mode to "Reciprocating" Mode or "Ultrasonic" mode or press
on the multifunctional footswitch (Fig. 4.1).
Notes
The E
C
™ will store the operative parameters for each "program" as soon as they are set by the user and they will be maintained in
NDO
ENTER
the memory even if the central unit is switched off.
To reset the parameters to the manufacturer presets (Software unit reset), press the "MODE" key while the unit is switched on . The LCD
screen will display the message: "SOFTWARE RESET"

5.2. SETTING IN ROTATING MODE

Torque Calibration:
1. Select "Rotating" using the MODE key or the multifunctional footswitch (Fig. 4.1).
2. Plug the "E-Type" contra-angle onto the micromotor: use only contra-angle that conforms to ISO 3964.
3. Press "Ratio", and with the "+" and "-" keys, set the reduction ratio according to the contra-angle you selected. Press again "Ratio"
to valid your choice.
4. Perform a torque calibration, using the "Calib." key on the front panel while holding the micromotor. This function chec ks the
handpiece mechanical inertia, in order to off-set this value. This operation will take a few seconds, the motor will run at a different
speed, and the LCD screen will display the following message:
Calibrating...
Please wait
Notes
Wether in rotating or reciprocating mode, the calibration should be performed at each change or lubrication of the contra-angle. Time of
Calibration: about 8 sec.
Do not perform the calibration if the motor cord is not connected the unit. In case of error, perform a "System Reset" as described in 5.1.
Do not perform any operation during calibration.
The file must not be engaged in any operation when the calibration is performed.
Programs:
The library of the E
C
NDO
ENTER
Once you have selected ROTATING using the "MODE" key or the multifunctional footswitch (Fig. 4.1), use the "+"and "-" keys to toggle from
P1 to P5.
For each "program" it is possible to set: Ratio, speed, torque, type of auto-reverse, direction FWD or REV and Flow rate.
Soft
3
Pump: 2
™ enables to store 5 programs in rotating mode.
10