the proper position. Only if a proper trigger mode is applied,
the waveform can be displayed steadily.
5. No Display Responses to the Push-down of Run/Stop.
Check whether Normal or Signal is chosen for Polarity in the
TRIG MODE menu and the trigger level exceeds the waveform
range.
If it is, make the trigger level is centered in the screen or set the
trigger mode as Auto. In addition, with the Autoset button
pressed, the setting above can be completed automatically.
6. The displaying of waveform seems match slow after
increasing AVERAGE value in Acqu Mode (see "How to Set
the Sampling/Display" on P45 ), or a longer duration is set in
the Persist in Display (see "Persist" on P46).
It's normal as the Oscilloscope is working hard on many more
data points.