Line
Type
(Time&V
Voltage
oltage
type)
Window
(Wave
Extension
zoom
mode)
Line
Source
Perform the following operation steps for the time and voltage
cursor measurement of the channel CH1:
1. Push Cursor to display the cursor menu.
2. In the right menu, select Source as CH1.
3. Press the first menu item in the right menu, select
Time&Voltage for Type, two blue dotted lines displayed
along the horizontal direction of the screen, two blue dotted
lines displayed along the vertical direction of the screen.
Cursor measure window at the left bottom of the screen
shows the cursor readout.
4. In the right menu, select Line Type as Time to make the
vertical cursors active. If the Line in the right menu is select
as a, turn the M knob to move line a to the right or left. If b is
selected, turn the M knob to move line b.
5. In the right menu, select Line Type as Voltage to make the
horizontal cursors active. Select Line in the right menu as a
Time
Makes the vertical cursors active.
Makes the horizontal cursors active.
Main
Measure in the main window.
Measure in the extension window.
Turn the M knob to move line a.
a
Turn the M knob to move line b.
b
ab
Two cursors are linked. Turn the M knob
to move the pair of cursors.
CH1
Display the channel to which the cursor
CH2
measurement will be applied.
71