Guidance and manufacturer's declaration – Electromagnetic interference emission
®
The SonicWeld Rx
is intended for use in an electromagnetic environment as specified below. The user
should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
according to
IEC 61000-4-2
Fast electrical
transients/bursts
according to
IEC 61000-4-4
Surge according to
IEC 61000-4-5
Voltage dips, short
power interruptions
and supply voltage
fluctuations
according to
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field according to
IEC 61000-4-8
Note
72
IEC 60601
test level
±6 kV contact
discharge
±8 kV air discharge
±2 kV for mains
power lines
±1 kV for input and
output lines
±1 kV differential-
mode voltage
±2 kV common-mode
voltage
<5% U
T
(>95% dip in U
)
T
for ½ period
40% U
T
(60% dip in U
)
T
for 5 periods
70% U
T
(30% dip in U
)
T
for 25 periods
<5% U
T
(>95% dip in U
)
T
for 5 seconds
3 A/m
U
is the a. c. mains voltage prior to application of the test level.
T
Instructions for Use / Technical Description
Compliance level
±6 kV contact
discharge
±8 kV air discharge
±2 kV for mains
power lines
±1 kV for input and
output lines
±1 kV differential-
mode voltage
±2 kV common-mode
voltage
5% U
T
(95% dip in U
)
T
for ½ period
40% U
T
(60% dip in U
)
T
for 5 periods
70% U
T
(30% dip in U
)
T
for 25 periods
5% U
T
(95% dip in U
)
T
for 5 seconds
3 A/m
Ultrasonic Unit SonicWeld Rx
Electromagnetic
environment – Guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user requires continued
operation during mains power
interruptions, it is
recommended to power the
device from an uninterruptible
power supply.
Power-frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
®