Instructions for Use / Technical Description
Ultrasonic Unit SonicWeld Rx
5.1.2
Handpiece, by Hand
Remove the sonotrode from the handpiece.
•
Disconnect the connecting cable with handpiece from the ultrasonic unit.
•
Clean the handpiece with a clean, lint-free cloth moistened with a commercially available
disinfectant approved for use with instruments and based on ethanol (50/50) or isopropanol
(70/30) (e.g. neodisher
NOTICE
Risk of damage due to improper handling!
The handpiece must not be immersed in a disinfecting or ultrasonic bath.
Do not use any acetone-containing disinfectants for cleaning and disinfecting the handpiece.
5.1.3
Device, by Hand
Clean the SonicWeld Rx
commercially available disinfectant approved for surface disinfection and based on ethanol
or methanol (e.g. neoform MED AF by Dr. Weigert).
WARNING
Danger of death by electric shock!
Ingress of liquids into the device must be avoided under any circumstances!
•
•
Before cleaning or disinfecting the SonicWeld Rx
from the power supply in all poles.
•
Should moisture have penetrated into the device, dry it after disconnecting it from the
supply.
V. 1.0
®
®
MediClean by Dr. Weigert).
®
ultrasonic unit with a clean, lint-free cloth moistened with a
®
, the device must be fully disconnected
61